Set the Stability and Response to Pouring mode (menu item
23
, See 11.2)
if faster response is required.
x
When the
[POWER]
key is pressed in the Formulation standby state, the power
standby state is initiated.
x
Formulation Mode setting can be cleared without menu operation, by pressing the
[POWER]
key for about three seconds.
x
See 11.3 for details of stability detection.
x
See 13.5 for details of Zero Range.
x
Cannot be used together with Pretaring (13.7).
x
Weighing unit (12.1) and minimum display (9.2) cannot be changed when
Formulation Mode is on.
x
When Formulation mode is ON, fully-automatic span calibration by PSC (10.3.2) or
Clock-CAL (10.3.3) is not performed. Weight symbol keeps blinking when span
calibration is necessary. Span calibration with the built-in calibration weight or external
weights can be performed between sets of measurements (Formulation stand-by).
x
Not more than one of these seven functions can be set at a time. Setting the second
one with the menu will automatically clear the previously set function;
Auto Print(13.3), Auto Zero(13.4), Peak Hold(14.3), Interval Timer(14.4), Add-on
mode(14.5), Animal Weighing mode(14.6), Formulation mode(14.7).
Содержание CU series
Страница 1: ...2 ...
Страница 11: ...12 2 Name and Function of Components 2 1 Components a Large pan model b Small pan model minimum display 0 01g ...
Страница 12: ...13 c Small pan model minimum display 0 001g windbreak standard a b c common ...
Страница 19: ...20 a Large pan model b Small pan model minimum display 0 01g ...
Страница 20: ...21 c Small pan model minimum display 0 001g ...
Страница 106: ...107 ...
Страница 107: ...108 ...
Страница 108: ...109 ...
Страница 109: ...110 ...
Страница 120: ...121 W Weighing 20 WindowsDirect 22 91 107 Z Zero range 60 Zero tracking 50 ...
Страница 124: ...125 MEMO ...
Страница 125: ...126 MEMO ...
Страница 126: ...127 ...
Страница 127: ...128 ...