115
A-6. Performance Checks
x
Performance checks should be conducted in a room where there are no sudden
temperature changes. Refer to the installation guidelines for the environmental factors
that assure optimal performance.
x
The following is a standard method used to determine whether the balance is working
properly.
x
The specific criterion should be set according to each user’s quality goal.
Repeatability
1
. Allow the balance to warm up sufficiently by
turning ON the power and leaving it at the
gramdisplay at least two hours before starting
the performance checks.
2
. Choose a weight that is near half the capacity
of the balance. Load and unload the weight five
successive times and record the following items:
Xi: Displayed value when the weight is loaded.
Yi: Displayed value when the weight is
unloaded.
3
. Use the formulas shown below to calculate the
values Rx and Ry.
4
. Balance repeatability is considered normal
when both Rx and Ry are within five counts.
* One count corresponds to one minimum
display unit of the balance model.
Содержание CU series
Страница 1: ...2 ...
Страница 11: ...12 2 Name and Function of Components 2 1 Components a Large pan model b Small pan model minimum display 0 01g ...
Страница 12: ...13 c Small pan model minimum display 0 001g windbreak standard a b c common ...
Страница 19: ...20 a Large pan model b Small pan model minimum display 0 01g ...
Страница 20: ...21 c Small pan model minimum display 0 001g ...
Страница 106: ...107 ...
Страница 107: ...108 ...
Страница 108: ...109 ...
Страница 109: ...110 ...
Страница 120: ...121 W Weighing 20 WindowsDirect 22 91 107 Z Zero range 60 Zero tracking 50 ...
Страница 124: ...125 MEMO ...
Страница 125: ...126 MEMO ...
Страница 126: ...127 ...
Страница 127: ...128 ...