CAS ELECTRICS CO.,LTD
4.1.8.3
Hysteresys Calibration
(Menu Code 8185)
(Calibration MENU -> 1. Calibration -> 8. Factory Setting -> 5. Hysteresys Calibration)
① Selecting “Hysteresys Calibration”
② For 5kg Mid-weight, input [5]key and press [ENTER]
③ Empty tray and press [ENTER]
④ While calibrating zero display shows “Wait4” ~ “Wait0” and follow next message for Mid Calibration.
⑤ Put on the Weight for Mid-weight then press [ENTER]
⑥ While calibrating Mid-weight display shows “Wait4” ~ “Wait0” and follow next message
for Span Calibration.