CAS ELECTRICS CO.,LTD
4.1.8.1 A/D Initialize (Menu Code 8183)
(Calibration MENU -> 1. Calibration -> 8. Factory Setting -> 3. A/D Initialize)
CAUTION: Must record setting values before Selecting [Y]. This will set the scale first default setting
4.1.8.2 Linearity Fine Adjust (Menu Code 8184)
(Calibration MENU -> 1. Calibration -> 8. Factory Setting -> 4. Linearity Fine Adjust)
① Selecting“LinearityFineAdjust”
ⓐ
Menu Code
ⓑ
Internal value
ⓒ
External value (weight)
ⓓ
Real value of Zero & Span
NOTE:
You can set 0 by pushing [ZERO]. This will update new Zero value.
② Put 5kg(Mid) on the tray.
③ Using cursor key for fine adjust.
* How to use cursor key
[▶] Increase Span value(ⓓ) to reduce internal (ⓑ)value
[◀] Decrease Span value(ⓓ) to increase internal (ⓑ)value
- Setting Mid value press [▼]key
- Internal value 20005 to change 20000 press [▶] 5times.
* Insert setting value by number key pad
ⓐ
ⓑ
ⓒ
ⓓ