4.1.2 Span/Zero Fine Adjust (Menu Code 8120)
(Calibration MENU -> 1. Calibration -> 2. Span/Zero Fine Adjust)
This mode is for fine tuning of scale after span Cal. Please put Max weight on the tray and adjust A/D
results at 60000, using the cursor key [◀ ▶.] and [Enter] key.
① Select menu “Span/Zero Fine Adjust”
ⓐ
Menu Code
ⓑ
Internal value
ⓒ
External value (weight)
ⓓ
Pure setting value of zero & Span
* If ⓒ is not set to zero press [ZERO] key. Value ⓓ will update.
② Put Max. Capacity weight on the tray
③ Use [◀] [▶] key for fine adjust.
* Insert setting value by cursor key (for the fine tune)
[▶] Increases Span value “ⓓ” to decrease Internal value “ⓑ”
[◀] Decreases Span value “ⓓ” to increase Internal value “ⓑ”
Ex) Internal value 60014 needs to change 60000
Press [▶] X 14 Times to decrease internal value.
ⓐ
ⓑ
ⓒ
ⓓ