83
(Operating Formulation Mode)
When the Formulation Mode is set on, the add-on
symbol and the Memory symbol are illuminated in
the mass display.
1
In the Formulation Mode, place the container
(if used) on the pan and press
[O/T]
key to tare.
(Taring is accepted only before weighing the
first sample.)
2
Place the sample (first component) in the
container and press
[PRINT]
key.
Upon stability, the mass value is outputted to the
external Device with the numbering “CMP001”.
After output, the display is automatically tared.
3
Repeat the above step 2 until all the component
samples have been weighed.
4
Press the
[POWER]
key.
The measurements up to this point are summed
up and displayed on the balance and outputted to
the external device.
5
Clear the pan.
The balance is ready for the next set of
measurements.
The items above are outputted for the
measurement examples shown here.
Decimal points can be either comma or
period in the output.(Refer to 14.4)
---Formulation Mode---
CMP001 =
0,5361g
CMP002 =
0,5422g
CMP003 =
0,4488g
TOTAL =
1,5271g