background image

 

41

7.4.3 Menu Lock 

 

The menu setting operations can be locked so that the settings cannot be inadvertently 
changed. This is called Menu Lock. Windows Direct settings also lock. 
The menu lock is set with the following procedure. 

 

How to lock the menu 

 

1

 Connect the balance to the power and wait. 

 
 

When “oFF” appears, press and hold the 

[CAL] 

key for about three seconds. 

“LoCKEd” appears and the menu lock 
function is activated, returning to the “oFF” 
display. When the lock function is activated,   
the MENU mark appears. 

 
 
 

When menu is locked, MENU symbol is 
illuminated at STAND-BY display, too.

 

 

 
 
 
 

If a menu item selection is attempted in locked status, the message “LoCKEd” appears and 
the menu selection is not allowed. To allow menu selections again, follow this procedure. 

 

How to remove menu lock 

 

1

 Disconnect the balance from power. Wait, 

then connect power again. 

 

2

 When “oFF” appears, press and hold the 

[CAL] key for about three seconds. 

 

3

 “rELEASE” appears and the menu lock is 

released.

 

 

(Time is not displayed with AUY.)

Содержание CAU Series

Страница 1: ...2 ...

Страница 2: ...Provides additional information needed to properly use the balance Other conventions used in this manual include Item Description 1 2 3 Indicates the step number in a procedure or a sequence of changes in the balance display key Indicates the operation key on the balance See 2 2 Indicates the message appearing in the balance display mass display Indicates that the balance is in the weighing mode a...

Страница 3: ...nts or where it may be splashed by water Ensure that the power supply voltage meets the indicated range of the AC adapter x Handle the balance carefully The balance is a precision instrument of solid design x Do not connect peripheral devices other than those recommended by CAS The balance may not operate properly if peripheral devices other than those specified in this manual are used The specifi...

Страница 4: ... Directives EMC directive 89 336 EEC amended by 92 31 EEC 93 68 EEC EN55022 1994 A1 1995 A2 1997 Class A EN55024 1998 EN61000 3 2 1995 A1 1998 A2 1998 EN61000 3 3 1995 Low Voltage directive 73 23 EEC amended by 93 68 EEC EN60950 1992 A1 1993 A2 1993 Weighing Instruments Department Analytical Measuring Instruments Division CAS CORPORATION CAS BLDG 440 1 SUNGNAE DONG GANGDONG GU SEOUL KOREA ...

Страница 5: ...6 ...

Страница 6: ... CAS CLASS BalanceAgent software part of a comprehensive laboratory data management system CAS CLASSAgent Ask your CAS representative about it CAS WindowsDirect When CAS balances are integrated with laboratory software by means of our WindowsDirect function no communication software is required or used The CAS balance functions as a primary device in the system just as a keyboard mouse or other da...

Страница 7: ...f the balance 25 5 1 Weighing 25 5 2 Changing the Unit Display 27 5 3 Switching the Weighing Range CAUW D series only 27 5 4 For Stable Measurement in Semi micro Range CAUW D series only 28 6 WindowsDirect Function 30 6 1 What is WindowsDirect 30 6 2 WindowsDirect Settings 30 6 2 1 Setting the Balance 30 6 2 2 Connecting the RS 232C Cable 31 6 2 3 Setting Up the Computer 32 6 2 4 Start and Checkin...

Страница 8: ...AUW D CAUW CAUXseriesonly 50 10 2 2 Span Calibration With External Weights 51 10 2 3CalibrationCheckWithBuilt inWeight CAUW D CAUW CAUXseriesonly 52 10 2 4 Calibration Check With External Weights 53 10 3 Calibration Settings 54 10 3 1 Selecting Preset Calibration Method 54 10 3 2 PSC Fully automaticspancalibration CAUW D CAUW CAUX series only 55 10 3 3Clock CALFully automaticspancalibration CAUW D...

Страница 9: ...on Functions 73 13 1 Piece Counting PCS 73 13 2 Solid Specific Gravity Measurement 74 13 3 Liquid Density Measurement 76 13 4Auto Print 78 13 5 Interval Timer CAUW D CAUW CAUX series only 79 13 6Add on Mode 80 13 7 Formulation Mode 82 14 Communication with Peripheral Devices 85 14 1 Electronic Printer 85 14 2 Personal Computer RS 232C 86 14 2 1 Connecting the Cable 86 14 2 2 Data Format 87 14 2 3 ...

Страница 10: ... 16 2 Troubleshooting 100 Appendices 101 A 1 Menu map 101 A 2 StandardAccessories and Maintenance Parts List 105 A 3 SpecialAccessories Options List 105 A 4 Specifications for RS 232C Connector 106 A 5 Table of Unit Conversion Constants 107 A 6 Performance Check Guide 108 A 7 Below weigh Hook Dimensions 110 A 8 Index 111 ...

Страница 11: ...nstruction manual before using the balance and keep the manual for future reference This manual has notation for CAUW D series CAUW series CAUX series CAUYseries combined as CAUWD CAUW CAUX CAUYseries or notations two to three series Note that the menu settings and the calibration display examples shown in this manual are mainly for the CAUW220 model On different models the numerical values and ot...

Страница 12: ... 1 Main Components Read these pages before installation Theft prevention ring For chains or other attachments Data I O connector DC IN connector RS 232C connector Below weigh hook cover Ground terminal attachment screw Keyboard connector ...

Страница 13: ...an electronic printer computer or other external devices Outputs the date and time to external devices Not with CAUY CAUW CAUX CAUY Switches display between 1d 10d Minimum display is rounded by one digit 1d 10d CAUW D Switches the weighing range During Menu Item Selection Key Pressed once and released Pressed and held for about 3 seconds POWER Return to the menu above the current menu level Return...

Страница 14: ...n the menu is locked Add on symbol Indicates set up ofAdd on mode or Formulation mode Memorysymbol Indicates set up of Formulation mode Communication symbol Indicates communication with external devices via RS 232C cable or Data I O connector Shown when communication functions are ON Batterysymbol Indicates a low batteryvoltage when using the balance with the optional batterypack Auto Print symbol...

Страница 15: ... 42grange 3 s Calibration weight Built in None Operation tem perature range 5 to 40 deg C Temperature coefficient of sensitivity 10 to 30 deg c 2ppm deg C when PSC function is Off 2ppm deg C Sensitivity stability 10 to 30 deg c 2ppm when PSC function is On Pan size 80mm diaApprox Main body dimensions 220mm W X 330mm D X 310mm H Main body Weight 7kgApprox Power requirements DC 10 to 15 5V 600mA Plu...

Страница 16: ...al accessory battery pack can be properly used Verify that the power voltage conforms to that indicated on theAC adapter 2 Installation site The balance must be installed in a proper place to ensure safe and accurate measurement Avoid the following types of sites Sites with air current from air conditioners ventilators open doors or windows Sites with vibration Sites in direct sunlight ...

Страница 17: ...top Stone is recommended Rather than the middle of the room the edges and corners are generally appropriate for vibration free measurement The glass doors open backwards beyond the rear end of the main body Make certain enough space is saved to allow the doors fully open 4 2 Unpacking and Delivery Inspection Analytical balance is a precision instrument Make certain not to allow any impact when pla...

Страница 18: ...cal distributor in case of damaged or missing items Standard packed items and quantity Standard packed items quantity Balance main body 1 Pan 1 Pan supporter 1 Anti draft ring 1 AC adapter 1 Adapter cable holder 1 Instruction manual 1 Explanatory operation sheet 1 Inspection certificate 1 In use protective cover 1 ...

Страница 19: ...tidraft ring 1 Gently attach the pan supporter on the center axis of the weighing chamber 2 Gently place the pan on the pan supporter 3 Place the anti draft ring 4Adjust the horizontal level level adjustment There are two level screws on the front of this balance The screws grow longer when turned in clockwise direction from above and they grow shorter when turned counter clockwise Adjust the leve...

Страница 20: ...21 5Attach the In use protective cover When the key panel and the display must be protected from dirt and wear place the cover over the key panel ...

Страница 21: ...by pressing the POWER key however at least one span calibration is necessary before use The CAUYseries does not perform calibration and the display showing CAL will not appear A power cable may be necessary to connect the AC adapter to the power source depending on the type of the AC adapter 3Press the POWER key After all displays appear the gram g display appears Also the display backlight illumi...

Страница 22: ...ically performed if necessary when the gram display first appears after the balance is turned ON See 4 4 When PSC operates the weight symbol starts blinking about two minutes beforehand for notification During operation the display will automatically change and the motor sound of the weight loading system is heard In order to ensure proper PSC operation prevent vibrations and air flow When gram di...

Страница 23: ...load the weights of the amount shown and close the glass door of the weighing chamber 5After a short time the zero display will blink Remove the weight from the pan and close the glass door After CALEnd appears the display returns to g display mode and the calibration is complete Span calibration by external weights is set for routine calibration by default in CAUYseries For other methods see 10 C...

Страница 24: ...ur hours 5 1 Weighing 1Open one of the glass doors of the weighing chamber place the weighing vessel container on the pan and close the glass door again When using a container 2Wait for the display to stabilize and press the O T key taring The appearance of the stability mark indicates a stable state The display will read zero 3Open the glass door place the items to be weighed in the weighing vess...

Страница 25: ...ep the glass doors closed unless otherwise described in this manual x Air convection causes measurement error if the temperatures of the weighed item and the chamber are different In order to avoid this equalize the temperatures by leaving the item in the extra space within the chamber before weighing ...

Страница 26: ... power is connected and turned on the balance is set to the small range with a minimum display of 0 01mg To switch to the large range with a minimum display of 0 1mg press the 1d 10d key except for CAUW D series this key has a different function t9 2 When measurement exceeds the small range capacity 82g for CAUW220D 42g for CAUW120D during use of the small range display automatically switches to t...

Страница 27: ...air conditioner When the air conditioner is in operation the temperature of the air current from it has a large difference from the room temperature Air current and presence of different temperatures both make the measurement unstable The temperature of the balance does not change as fast as the room temperature Changing room temperature prior to measurement causes temperature difference remaining...

Страница 28: ...e temperature of the samples to the chamber s Remove influence of the heat emitted from human body and the heat from the sample to be weighed Equalize the temperature of the object to be weighed to that of balance weighing chamber If possible keep the object in the extra space of the chamber prior to measurement Use a long pair of tweezers spatula or other instruments to keep hands away from the w...

Страница 29: ...odes 14 2 3 is required Only numerical values can be transferred through WindowsDirect 6 2 WindowsDirect Settings Simple settings are made for the balance and the computer Connection is by RS 232C cable specified by CAS If bi directional communication software is used WindowsDirect function should be turned off Set up the optimal communication parameters for the software according to 14 3 Communic...

Страница 30: ...Etwin or SEtwin setting individual communication parameters can be changed using the Communications Settings menu 14 3 In this case the stability mark may still appear if you go to the SEtwin or SEtwin display but WindowsDirect may not operate To restore WindowsDirect optimal settings then first remove the stability mark by pressing the O T key at the SEtwin or SEtwin display This restores the def...

Страница 31: ...01 6 2 3 Setting Up the Computer leave the balance unplugged 1 Turn ON the power to the computer and start Windows 2Click Start choose Settings and Control Panel 3Select Accessibility Options 4Verify that there are no check marks for any items on all five tabs including General 1 2 D sub9 1 2 3 4 5 6 7 8 9 D sub25 1 2 3 4 5 6 7 8 20 22 D sub9 2 3 4 5 6 7 8 D sub25 2 3 4 5 6 7 20 ...

Страница 32: ... options appears in the General tab Put check marks at both the items of Administrative options to maintain the settings even after restarting Windows 6Open Settings 7 Select the serial port corresponding to the RS 232C port of your personal computer Serial port any one of COM1 to 4 Usually COM1 8Select a Baud rate of 300 9Click OK 10Click Apply and wait 11Click OK ...

Страница 33: ...he application you wish to use 4Press the PRINT key o f the balance Verify that the numeric value displayed on the balance appears at the cursor position on the screen of computer The effect is the same as typing the value from the computer keyboard and pressing the ENTER key Characteristics indicating the unit of measure are not sent to the computer 5 Test combination withAuto Print function if y...

Страница 34: ... a check mark on Accessibility Options For more information see the Windows instruction manual When Support Serialkey device is selected in Accessability Options software which uses the same RS 232C port on that computer does not operate correctly until Serialkey support is discontinued If another device an external modem plotter or etc is to be connected remove the check mark placed on Support Se...

Страница 35: ...confirm that no data is entering the computer before touching the keyboard or the mouse x This function may generate incorrect data when the displayed value is not a weight value i e error code or time x The unit designation is not transmitted The balance display unit selected and the unit required by the application should be set the same x This function may operate incorrectly depending on the s...

Страница 36: ...ded in the CAUW D CAUW CAUX CAUY series Procedures of each menu item selection are explained with the display examples in chapters 8 to 14 However when selecting menu items refer to the menu map for more efficient setting 7 2 What is a menu map The CAUW D CAUW CAUX CAUYseries menu consists of four levels The menu map displays this hierarchy in an easy to understand format The map allows quick acce...

Страница 37: ...and released Pressed and held for about 3 seconds Moving direction on Menu Map POWER Return to the menu above the current menu level Returns to mass display CAL Moves to the next menu item O T Select the displayed menu item or move to the next menu below the current menu level UNIT In numerical value input increases the numeric value of the blinking digit by 1 PRINT In numerical value input moves ...

Страница 38: ...pecific gravity measurement see 13 2 13 3 and settings related to the clock Communication settings menu group Select This second level menu group includes settings for communication formats for connecting the balance to external devices see 14 3 There are standard formats prepared beforehand and user defined formats that allow itemized selection Mass display Execute the preset calibration method s...

Страница 39: ... either or both of the fully automatic span calibration PSC or Clock CALare on 7 4 2 Returning to Default Settings Menu reset This will return all the settings to default The reference value stored in previous use of piece counting or percentage conversion and the set time of Interval Timer function will also be cleared The default settings are indicated with on the Menu Map 1Press the CAL key rep...

Страница 40: ...k function is activated returning to the oFF display When the lock function is activated the MENU mark appears When menu is locked MENU symbol is illuminated at STAND BYdisplay too If a menu item selection is attempted in locked status the message LoCKEd appears and the menu selection is not allowed To allow menu selections again follow this procedure How to remove menu lock 1Disconnect the balanc...

Страница 41: ...nd press the O T key In the upper part of the display the MENU Symbol and the symbol appear indicating number entry mode Also the current date setting appears as YY MM DD YYis the year with the leftmost digit blinking 3Pressing the UNIT key increases the blinking digit by one Pressing the PRINT key moves the blinking digit one place to the right 4Press the O T key when the desired date is shown in...

Страница 42: ... shows y m d After this pressing the CAL key cycles the displays in the order of y m d m d y d m y y m d y m d is for settingYYYY MM DD m d y is for MM DD YYYYand d m y is for DD MM YYYY When the current setting is displayed the stability mark appears To change the setting press the O T key when the desired setting is on the display or 4Press the POWER key to return to StyL dAtE without changing t...

Страница 43: ...EMU symbol and the symbol appear indicating number entry status Also the current time setting appears as HH MM SS HH is hours MM is minutes and SS is seconds with the leftmost digit blinking 4Pressing the UNIT key increases the blinking digit by one Pressing the PRINT key moves the blinking digit one place to the right 5Press the O T key when the desired time is shown in the display The built in c...

Страница 44: ... on or oF 3Press the O T key Ad on is displayed At this point pressing the CAL key switches the display between Ad on and Ad oF When the current setting is displayed the stability mark appears To change the setting press the O T key when the desired setting appears and the display will return to AdiSP indicating the new setting 4If no change is required press the POWER key The display returns to A...

Страница 45: ...play The minimum display digit will be displayed with only three decimal places 2 To return the minimum display to the standard setting Press the 1d 10d key from the mass display The minimum display digit will return to the previous setting x The decimal point does not move When the minimum display is changed by one decimal place the final decimal place remains empty x In the CAUW D series the 1d ...

Страница 46: ...that point every time the CAL key is pressed the display switches between LiGt on on and LiGt oF off Here when the current setting is displayed the stability mark appears 5Press the O T key when the desired setting appears and the display will return to LiGHt indicating the new setting 6If no change is required press the POWER key to return to LiGHt 7Press the POWER key repeatedly to return to the...

Страница 47: ... This allows the mass of any item that is subsequently weighed to be computed from its weight and accurately displayed The recording and setting of weight mass relation is known as span calibration The necessity can be recognized this way for example the gravitational acceleration in the Japanese cities of Kyoto and Tokyo are 979 70775 cm s2 and 979 76319 cm s2 respectively with a difference of ab...

Страница 48: ...libration made necessary by temperature changes and Clock CAL for performing automatic span calibration at set times Calibration can also be done with external calibration weights E CAL E tESt Calibration of the built in calibration weight with external calibration weights PCAL is also possible Calibration of the CAUYseries must be done with external calibration weights as this series does not hav...

Страница 49: ...isplay shows i CAL Preset calibration method is i CAL jump to Step 5 If the display shows another item Preset calibration method is not i CAL go on to Step 2 Refer to 10 3 1 for preset calibration method 2Press the CAL key repeatedly until FUnC SEL appears then press the O T key to display CAL 3Press the O T key again The display shows E CAL 4Press the CAL key twice The display shows i CAL 5Press ...

Страница 50: ...nC SEL appears then press the O T key to display CAL 3Press the O T key The display shows E CAL 4While E CAL is shown press the O T key Span calibration begins and the zero display blinks 5Recheck that there is nothing on the pan and wait 6The set weight value Refer to 10 3 5 appears blinking 7Place the weight indicated on the pan Wait until the blinking zero display reappears It may take about 30...

Страница 51: ...ey repeatedly until FUnC SEL appears then press the O T key to display CAL 3Press the O T key again The display shows E CAL 4Press the CAL key repeatedly until the display shows i tESt 5Press the O T key The display changes automatically in this order tESt 2 tESt 1 and tESt 0 After that the display shows the drift from the previous span calibration for several seconds 6After tEStEnd appears for se...

Страница 52: ...dly until the display shows E tESt 5Press the O T key The calibration test begins and the zero display blinks 6Recheck that there is nothing on the pan and wait 7The set weight value Refer to 10 3 5 appears blinking 8Place the weight indicated on the pan Wait until the blinking zero display reappears It may take about 30 seconds 9Remove the weight The display shows the drift from the previous span...

Страница 53: ...peatedly until the SEttinG menu appears Press the O T key CALdEF appears 2When CALdEF appears press the O T key The E CAL display appear 3Press the CAL key repeatedly until the desired setting appears Select from the following four types or two types for CAUYseries The stability mark appears when the currently preset method is shown E CAL Span Calibration with external weights E CAL E tESt Calibra...

Страница 54: ...previous calibration 3 After either condition 1 or 2 above has been met during warm up at stand by status when the balance is switched to mass display mode In mass display mode when any of the conditions above has been met the weight symbol will blink for about two minutes as notification of span calibration before calibration begins The sensitivity before and after span calibration is slightly di...

Страница 55: ...PSC on Here when the current setting is displayed the stability mark appears When the desired setting appears pressing the O T key changes the current setting Pressing the POWER key returns the setting to PSC without changing the setting 5Press the POWER key repeatedly to return to the mass display PSC and Clock CALcan each be turned on and off independently Weight symbol appears in the Settings C...

Страница 56: ...ration Never cause vibration of the balance during calibration Setting PSC ON OFF 1From mass display press the CAL key repeatedly until SEttinG appears Press the O T key to display CALdEF 2Press the CAL key repeatedly until tCAL appears The positions show the current setting either on when on or oF when off 3To change the setting press the O T key when tCAL appears 4 tCAL on appears and from this ...

Страница 57: ...ange the value of the blinking digit When the UNIT key is pressed the value of the blinking digit increases by 1 at a time The numerals progress in this order 0 1 2 9 _ 0 Set the hours in the range from 00 to 23 and the minutes from 00 to 59 Press the POWER key to complete the setting The display returns to t CALt 4To set another time press the CAL key to move to the next t CALt setting and set th...

Страница 58: ...ing is on the pan and ensure conditions free from the influence of vibration or air flow Setting PSC ON OFF 1 From mass display press the CAL key repeatedly until SEttinG appears Press the O T key CALdEF appears 2 Press the CAL key repeatedly until PcAL appears and press the O T key Calibration of the built in calibration weight begins and the display shows PCAL the C becomes capital 3 When the di...

Страница 59: ... symbol appear in order to indicate numerical input mode The leftmost numeral blinks The blinking digit may be changed 3 When the UNIT key is pressed the value of the blinking digit increases by 1 at a time Press the PRINT key to move the blinking digit one place to the right Enter the desired value 4 Press the O T key to set the external calibration weight value The display proceeds to SEt and th...

Страница 60: ... in order to show number entry mode The leftmost numeral blinks The blinking digit may be changed 3 When the UNIT key is pressed the value of the blinking digit increases by 1 at a time Press the PRINT key to move the blinking digit one place to the right Enter the desired value 4 Press the O T key to set the external calibration weight value for built in weight calibration The display proceeds to...

Страница 61: ...ey repeatedly until SEttinG appears Press the O T key CALdEF appears 2 Press the CAL key repeatedly until GLP appears The positions show the current setting on when set and oF when unset 3 To change the setting press the O T key when GLP is showing The display will show GLPon The becomes 4 From this point every time the CAL key is pressed the display will switch between GLP oF and GLP on Here when...

Страница 62: ... is a number Press the O T key In the upper part of the display panel the MENU symbol and the symbol appear in order to indicate numerical input mode The leftmost digit of blinks The numeral of the blinking digit can be changed 3 Press the UNIT key to increase the value of the blinking digit by 1 Press the PRINT key to set that digit and move the blinking digit one place to the right When the desi...

Страница 63: ...nt 3 To change the setting press the O T key when Prtdt is showing The display will show Prdt on The becomes 4 From this point every time the CAL key is pressed the display will switch between Prdt oF and Prdt on Here when the current setting is displayed the stability mark appears 5 To change the setting press the O T key when the desired setting appears or 6 Press the POWER key to return to Prtd...

Страница 64: ...the objectives of using the balance Anti convection mode High stability mode and Pouring mode are also available The currently set mode is easily checked by pressing the CAL key four times from mass display to produce the settings check display see 7 4 1 11 2 1 Standard mode This is the default setting Use this mode unless stability or response is affected by environmental factors From the mass di...

Страница 65: ...s displayed Press the O T key here to enter High stability mode Setting of this mode can be confirmed only in settings check display see 7 4 1 11 2 4 Pouring Mode fast response When using the balance for dosing or filling a specified quantity Pouring mode will give sufficiently fast response Note that the display will become very sensitive and unstable Pouring mode allows adjustments to the surrou...

Страница 66: ...r UnStb Env when you find the measurements are too slow or too unstable with the default normL Env Settings check display Setting of Pouring mode can be confirmed only in the settings check display see 7 4 1 Settings check display also shows the environmental setting for Pouring mode when this mode is selected when normal is selected Pouring mode for stable environmental condition Pouring mode for...

Страница 67: ...ing the Stability Detection Band 1 Press the CAL key repeatedly from the mass display When FUnC SEL appears press the O T key CAL is displayed 2 Press the CAL key twice bAnd is shown The part single or double digits shows the current stability detection band setting 3 Press the O T key b 1 appears At this point pressing the CAL key changes the display in this order b 1 b 5 b 10 representing stabil...

Страница 68: ...splay CAL 2 Press the CAL key once trC is shown The part shows the current setting as on for on and oF for off 3 Press the O T key trC on appears The changes to At this point pressing the CAL key changes the display between trC oF zero tracking off and trC on zero tracking on When the current setting is displayed the stability mark appears 4 Press the O T key to select the zero tracking setting cu...

Страница 69: ...ounting and specific gravity measurements are also covered in section 12 1 For information about piece counting refer to 13 1 Refer to 13 2 for solid specific gravity measurement and 13 3 for liquid density measurement 12 1 Setting Units of Measurement 1 Press the CAL key repeatedly from the mass display until FUnC SEL is showing Press the O T key to display CAL 2 Press the CAL key repeatedly unti...

Страница 70: ...ngapore tael U tiwAn Taiwan tael U mAL Malaysia tael U ChinA Chinese tael U dwt Pennyweight U GN Grain U m Mesghal U b Baht U t Tola U o Parts Pound Depending on the legal restriction these units are not always available 3 Select the units to register by pressing the O T key when the desired unit appears To remove a registered unit press the O T key when the unit to be removed appears with stabili...

Страница 71: ...am display and wait for the stability mark to appear 5 Press the CAL key After SEt is displayed the mass of the standard item is set as 100 6 Remove the standard item and begin measurement by loading samples 7 To return to the mass display press the UNIT key several times x When the UNIT key is pressed the display returns to the mass display with units such as g The actual mass of the current samp...

Страница 72: ...s order Ld 10 Ld 20 Ld 50 and Ld 100 7 When the correct number of pieces on the balance is displayed and the stability mark has appeared press the O T key 8 SEt appears for several seconds and the number of pieces is displayed 9 Load the items to be measured and read the number of pieces x When the UNIT key is pressed the display returns to the mass display with units such as g The actual mass of ...

Страница 73: ... d the solid specific gravity as a unit referring to 12 1 4 Input the density of the liquid to be used for solid specific gravity measurement a Press the CAL key repeatedly from the mass display until the SEttinG appears Press the O T key The display shows CALdEF b Press the CAL key repeatedly until LSG SEt appears and press the O T key MENU The symbol and the symbol appearing in the upper part of...

Страница 74: ... pan The display shows the density of the measured item dSPoL may appear when nothing is on the pan which is normal 10 To make the next measurement unload the pans press CAL and begin again at Step 5 x Although the density display shows 4 decimal places not all of these digits will stabilize depending on the conditions x When the items to be measured are placed in the liquid they must be completel...

Страница 75: ...d density as a unit referring to 12 1 4 Input the sinker volume in cm3 for liquid density measurement a Press the CAL key repeatedly from the mass display until the SEttinG display appears Press the O T key The display shows CALdEF b Press the CAL key repeatedly until Sv SEt appears and press the O T key appears where the is a number The MENU symbol and the symbol appearing in the upper part of th...

Страница 76: ...nging pan The display shows the density of the sample dSPoL may appear when nothing is on the pan which is normal 10 To make the next measurement unload the pans press CAL and begin again at Step 5 x Although the density display shows 4 decimal places not all of these digits will stabilize depending on the conditions x When the sinker is placed in the sample liquid it must be completely immersed a...

Страница 77: ...isplay will show CAL 2 Press the CAL key repeatedly until AtPrt appears the shows the current setting on for on and oF for off 3 Press the O T key The display shows AP on and after this pressing the CAL key switches the displays between AP on and AP oF When the current setting is displayed the stability mark appears To change the setting press the O T key when the desired setting appears or 4 Pres...

Страница 78: ...ost digit blinks 4 Pressing the UNIT key increases the blinking digit by 1 Pressing the PRINT key moves the blinking digit one place to the right Set minutes up to 99 and seconds up to 59 5 Press O T key to set 6 Press the POWER key repeatedly to return to the mass display The letter T and STAND BY symbol appear indicating interval output standby 7 Press the PRINT key Automatic output commences an...

Страница 79: ...while using interval timer function 13 6 Add on Mode This function is convenient for making many measurements of minute samples In this mode with any unit when the displayed value is within 5 counts of zero and an item of mass of 10 counts or more is placed on the pan that value is output via the RS 232C cable or DATAI O connector upon display stability and automatic taring is done each time after...

Страница 80: ... when the desired setting is shown to change the setting 4 Press the POWER key to return to Addon 5 Pressing the POWER key again returns to the mass display 6 When theAdd on Mode is set up the add on mode symbol appears in the mass display x No more than one of the four application modes Auto Print 13 4 Interval Timer 13 5 Addon Mode 13 6 Formulation Mode 13 7 can be set ON at the same time x When...

Страница 81: ...ing Formulation Mode 1 Press the CAL key repeatedly from the mass display until FUnC SEL appears Press the O T key The display will show CAL 2 Press the CAL key repeatedly until Formu appears The shows the current setting on for on oF for off 3 Press the O T key The display shows Formu on and after this pressing the CAL key switches the displays between Formu on and Formu oF When the current setti...

Страница 82: ...tted to the external Device with the numbering CMP001 After output the display is automatically tared 3 Repeat the above step 2 until all the component samples have been weighed 4 Press the POWER key The measurements up to this point are summed up and displayed on the balance and outputted to the external device 5 Clear the pan The balance is ready for the next set of measurements The items above ...

Страница 83: ...s in use fully automatic span calibration by PSC 10 3 2 or Clock CAL 10 3 3 is not performed Instead Weight symbol keeps blinking when span calibration is necessary Span calibration can be performed between sets of measurements ...

Страница 84: ...lance is either oFF or STAND BY Then connect the printer cable to the DATAI O on the back of the balance Connect the printer cable to the printer too 4 Turn on the balance power 5 Turn on the printer power x Refer to the instruction manual of the electronic printer x When using EP 80 or EP 50WIN simultaneously with WindowsDirect function communication settings are different Follow 6 2 1 for settin...

Страница 85: ...cable must have the correct connections as shown in the diagram below x Cables with the connections shown below and the optional accessory RS 232 cable are not guaranteed to operate properly for all types of computers and devices x Refer to 6 2 2 when using WindowsDirect function For DOS V computers D sub9 pin Null modem Computer side Balance side RXD 2 2 TXD TXD 3 3 RXD DTR 4 6 DSR SG 5 7 SG DSR ...

Страница 86: ...ositions the code for space 20H is entered to each excessive position as shown in this example With CAUW D space may appear after the value at the position No 10 Positions 11 12 One or two letters indicating the weighing unit As shown in this example the code for space is entered at position No 12 if only one letter is used for the unit Position 13 Code for delimiter 2 Information of additional by...

Страница 87: ...in case of oL or oL Overload The below is the data format for oL Data length of this example 13 bytes Position 1 2 3 4 5 6 7 8 9 10 11 12 13 ASCII code 20H 20H 20H 20H 20H 4FH 4CH 20H 20H 20H 20H 20H 0DH Data O L C R For oL negative overload Position 1 is replaced with minus ASCII code 2DH The following parts appearing in 2 Information of additional bytes also apply to 3 i Output with stability in...

Страница 88: ...tatus of the stability mark is appended to the head of the data with output S when the stability mark is showing U when the mark is not showing D08 1 time output at stability After command input the data is output at the first appearance of the stability mark D09 Halt output Auto Print and continuous output halted Q ON OFF switching Switches between standby status and measurement status T Taring C...

Страница 89: ...Solidspecificgravityremoval LDENCE Liquidspecificgravityremoval C02 High stabilitymodesetting C13 Anti convectionmodesetting C14 Standardmodesetting C05 Stabilitydetectionband 1countsetting C06 Stabilitydetectionband 5countsetting C15 Stabilitydetectionband 10countsetting C07 Zerotracking C08 Unsetzerotracking C10 Automaticspancalibration C11 Unsetautomaticspancalibration C17 Display setting statu...

Страница 90: ...items in the User Settings 14 3 2 Standard Setting The combinations shown in the table below are available as standard setting 1 to standard setting 6 In the communication settings menu the standard settings 5 and 6 cannot be selected The standard settings 5 and 6 WindowsDirect can be set easily without entering the menu see 6 2 1 Displayat menu item selection Compatible manufacturer Baud rate Del...

Страница 91: ...s display press the CAL key repeatedly until intFACE appears and press the O T key iF F1 appears 2 Press the CAL key repeatedly until iF USEr appears and press the O T key io b appears communication speed settings shows the current setting 2 to 4 characters the same applies hereafter After this pressing the CAL key alternates the display in this order io d delimiter setting io P parity setting io ...

Страница 92: ...he display changes from io d to d Cr Pressing the CAL key cycles through the available settings The stability mark appears when the current setting is displayed Display during setting d Cr d LF d CrLF d Cn d win Setting specifics delimiter C R delimiter L F delimiter C R L F delimiter Comma delimiter WindowsDirect 2 When the desired setting appears press the O T key 3 Press the POWER key to return...

Страница 93: ...ssing the CAL key cycles through the available settings The stability mark appears when the current setting is displayed Display during setting F dF1 F dF2 F dF3 F dF4 Setting specifics Data format 1 Standard CAS format Data format 2 Extended format from data format 1 Data format 3 Same format as Mettler balances Data format 4 Same format as Sartorius balances 2 When the desired setting appears pr...

Страница 94: ...Note that the decimal point expression on the balance display is always with period 1 From the mass display press the CAL key repeatedly until SEttinG appears press the O T key CALdEF will appear 2 Press the CAL key repeatedly until dECPt appears indicates the current setting Pr for period Co for comma 3 Press the O T key The display shows dECPt Co and after this pressing the CAL key switches the ...

Страница 95: ...he display panel Handle the glass doors with great care to keep them from breaking When removing the knob on the inside of the door be very careful not to allow the hand to touch the pan supporter shaft in the weighing chamber When removing the door rails take care not to let the rail edge injure the hands When the glass door does not slide smoothly x In the CAUW D CAUW CAUX CAUYseries the side gl...

Страница 96: ... when the door rails of the side glass doors become dirty or worn the rails can be replaced Removing the door rail 1 Remove the glass door 2 Press down the outer edge of the door rail with a pointed thing for lifting the door rail 3 Lift and remove the rail Door rail Door rail ...

Страница 97: ...in the figures and carry it securely in both hands When using other methods of transport Use the packing box in which the balance was delivered 1 Place hands on the main body as shown in the figure 2 Lift the main body slowly with the fingers inserted beneath the bottom of it This is a precision instrument Handle with care and never give any impact ...

Страница 98: ...PCAL Use the correct calibration weight CALE4 Large span error in span calibration Use the correct calibration weight CHE X X is a numeral when the display stops here Internal malfunction Please contact a service representative Err 0X X is a numeral Internal malfunction Please contact a service representative Err 20 An improper value setting was attempted Enter the correct numbers or decimal point...

Страница 99: ...sample temperature and the temperature inside the weighing chamber differ Measure at the same temperature Leave the item in the chamber before measurement Change to high stability mode There are air currents entering the weighing chamber Leave the glass doors of the weighing chamber open 1 to 2 cm wide when not in use Effects of electronic noise or strong electromagnetic waves Move away from the n...

Страница 100: ...essage appears Refer to the error code table During calibration Error message appears Refer to the error code table Attempt of menu settings The menu cannot be entered LoCKEd is displayed briefly The menu is locked Remove the menu lock See 7 4 3 Appendices A 1 Menu map in brackets after the menu item shows reference section number x CAL key Pressing the CAL key moves to the next menu in the curren...

Страница 101: ... Mesghal unit Preset delete Baht unit Preset delete Tola unit Preset delete Parts pound unit 0 0000 i CAL Ex Mass Display WindowsDirect Setting Down 6 2 1 WindowsDirect Setting Right 6 2 1 Standard mode 11 2 1 Anti convection mode 11 2 2 High stability mode 11 2 3 Mass Display Excuting preset calibration method 10 2 1 Displays the preset calibration method Mass Display Mass Display St b1 t Ex Disp...

Страница 102: ...ght 9 3 Backlight on Backlight off CAL SEt PCAL SEt 1 PcAL 1 LigHt on 2 Clock CAL running time setting 10 3 3 Clock CAL running time setting 10 3 3 Clock CAL running time setting 10 3 3 tCAL t1 3 tCAL t2 3 tCAL t3 3 Setting of solvent specific gravity for solid specific gravity measurementent 13 2 Setting of sinker volume for liquid density measurement 13 3 LSG SEt Sv SEt Date setting 8 1 Date out...

Страница 103: ...Cn d win d win Delimiter setting io F dF1 F dF1 F dF2 F dF3 F dF4 Data format setting io H oFF H oFF H SoFt H HArd H tr Handshake setting io P no P no P odd P EvEn Parity setting io S S1 S S1 S S2 C R L F C R L F Comma WindowsDirect down WindowsDirect Right Data format 1 Data format 2 Data format 3 Data format 4 None Software Hardware Timer No parity 8 bits Odd parity 7 bits Even parity 7 bits 1 b...

Страница 104: ... 3 SpecialAccessories Options List Part name Part number Remarks Electronic Printer EP 80 321 62675 01 w oAC adapter Impact dot print Can be used with WindowsDirect Statistical calculation Electronic Printer EP 90 321 62675 11 w oAC adapter Impact dot print Can be used with WindowsDirect Statistical calculation Sample No can be attached Date time can be attached except for CAUY Ionizer Static Remo...

Страница 105: ...S 232C Connector RS 232 Specifications Pin number Name Function Comment 1 FG Frame ground 2 TXD Data output 3 RXD Data input 4 RTS Internal connection with CTS 5 CTS Internal connection with RTS 6 DSR Handshake reception 7 SG Signal ground 8 NC Open 9 NC Open 10 NC Open 11 NC Open 12 NC Open 13 NC Open 14 NC Open 15 NC Open 16 NC Open 17 NC Open 18 NC Open 19 NC Open 20 DTR Handshake transmission ...

Страница 106: ...nits 1g 1000 mg 5 ct 0 266667 mom 0 00220462 Lb 0 0352740 Oz 0 0321507 Ozt 0 0267173 TL HK 0 0264555 TL Singapore 0 0266667 TL Taiwan 0 0264600 TL Malaysia 0 0266071 TL China 0 643015 dwt 15 4324 GN 0 216999 m 0 0657895 b 0 0857339 t 1 12877 o The unit of parts pounds o is not outputted to external devices ...

Страница 107: ...r Repeatability For CAUW D series perform steps 3 to 5 in both large and small ranges 1 Leave the power on in mass display backlight of CAUW series illuminates if it is set to on and let stand for one hour two hours for CAUW D series 2 Following chapter 11 set the stability and response to the standard mode set the stability detection band to 1 count and set zero tracking to OFF 3 With a pair of t...

Страница 108: ... about one quarter of the capacity on the pan at the locations and in the order of the numbers given in the figure Record the results X1 through X5 for the locations 1 through 5 respectively If all the differences between the value measured in the center location and each of the other locations are within 1mg the balance cornerload performance is considered normal In CAUW D series this check is no...

Страница 109: ...110 sensor Below weigh hook 2 Below weigh hook Balance bottom H o l e 2 0 1 5 1 5 3 6 3 5 2 5 16 5 A 7 Below weigh Hook Dimensions Unit mm ...

Страница 110: ...alibration weight 54 calibration test with external weights 54 calibration weight value 60 61 Clock CAL 23 49 57 62 command codes 30 89 communication setting 30 91 communication speed 36 91 92 component names 13 computer 30 32 85 86 91 continuous output 89 current time 22 44 D data format 87 date 42 43 64 DC IN connector 13 22 decimal point 46 95 default settings 30 40 102 103 104 delimiter 91 93 ...

Страница 111: ...uter personal computer 30 86 piece counting 73 pouring mode 66 power 22 31 85 preset calibration method 54 printer electronic printer PSC 23 55 R range weighing range response 65 RS 232C 13 31 86 106 RS 232C conector specification specifica tions for RS 232C connector S safety precaution 12 self check 22 setting calibration type of regular use 54 setting units 70 settings check display 39 40 66 sm...

Страница 112: ...indowsDirect function 35 types of calibration 49 U unit 27 70 unit conversion 107 unit display 27 Unpacking 17 user settings 91 92 V vibration 17 23 65 66 W warm up 22 23 25 50 55 59 109 warm up 47 weighing chamber 13 20 25 96 weighing range 14 27 WindowsDirect 30 78 86 Z zero tracking 69 ...

Страница 113: ...ronic printer or a computer PRINT To output date and time Press and hold forabout 3 seconds Not with CAUY CAUW CAUX CAUY Switches minimum display 1d 10d CAUW D Switches weighing range Stabilizing Display x If it is unavoidable to install the balance at a location subject to air flow or vibration try changing the environmental settings to high stability mode How to set high stability mode From mass...

Страница 114: ...ow change in mass switch Zero Tracking OFF How to switch off Zero Tracking From mass display Press CAL key repeatedly till FunC SEL is displayed Press O T key CAL is displayed Press CAL key once trC on is displayed Press O T key trC on is displayed Press CAL key trC oF is displayed Press O T key is added Press POWER key trC oF is displayed Press POWER key repeatedly to return to the mass display S...

Страница 115: ...From mass display press CAL key repeatedly until SEttinG appears Press O T key to display CAL dEF Press CAL key repeatedly until PSC on appears Press O T key PSC on appears Press CAL key PSC oF appears Press O T key Stability mark is added Press POWER key returns PSC oF Press POWER key repeatedly to return to the mass display When PSC is off perform span calibration according to Span Calibration P...

Страница 116: ... CAL2 CAL1 CAL0 and CALEnd it returns to the mass display and span calibration is complete When fully automatic span calibration PSC is on span calibration is automatically performed according to temperature changes PSC is set to ON by default See 10 3 2 CAUYSeries Span Calibration Using External Weights 1 PressFrom the mass display press the CAL key The display shows E CAL default settings 2 Pres...

Страница 117: ...118 MEMO ...

Страница 118: ...119 ...

Страница 119: ...120 ...

Отзывы: