Mount the recorder by using the already installed bracket.
There are three factory drilled holes located on the right side of chamber facing the front of
chamber.
Using the factory supplied screws, screw the recorder to the side of the chamber.
Model 6030, 6031, 6032
Right side
Model 6010, 6011, 6012
Right side
NOTE: Before applying power to the recorder, complete these preliminary preparation
tasks to prepare the recorder for operation.
NEVER
ATTEMPT
TO
ACCESS
COMPONENTS
INSIDE
THE
RECORDER
CASE
WITH
POWER
APPLIED.
The Vaisala RH Sensor is wired to the Humidity controller in the chamber for 0-5 Volt actuation.
An individual 100
Ω
RTD Temperature sensor is wired to the Recorder.
In order to connect the Humidity and temperature sensor to the recorder, refer to the figures below
to perform the following procedures: