CS 7200 User and Installation Guide (6M0940)_Ed04
43
4. Wait for the Scanner Status icon to turn green on the
Imaging Window
toolbar indicating that the CS
7200 scanner is connected to the workstation and is ready for acquisition.
Preparing the Imaging Plates
To prepare the intraoral imaging plate, follow these steps:
1
Choose an imaging plate of the appropriate size for your examination.
2
Insert the imaging plate into its protective hygienic sheath with the inactive side facing the adhesive
strip, so the imaging plate’s size number is seen through the transparent side of the hygienic sheath and
the orientation mark visible in the bottom corner (
A
).
Important: Use only CS 7200 intraoral imaging plates from Carestream
and hygienic sheaths from Carestream. The use of plates or hygienic
sheaths from a third party supplier may cause a malfunction of the
system and void the warranty
.
Important: Use a NEW hygienic sheath for each new patient to prevent
cross contamination.
Important: Use a NEW hygienic sheath for each new image to prevent
humidity and contamination of the plate.
Tip:
If the imaging plate has not been used for a while, proceed to erasing it
first. See
Содержание CS 7200
Страница 1: ...User and Installation Guide CS 7200...
Страница 6: ...2 Chapter 1 Conventions in this Guide...
Страница 31: ...CS 7200 User Installation Guide 6M0940 _Ed04 27 11 Click Finish to finish the installation process...
Страница 38: ...34 Chapter 3 Setting Up CS 7200...
Страница 66: ...62 Chapter 5 Acquiring Images...
Страница 68: ...64 Chapter 6 Maintenance...
Страница 80: ...76 Chapter 8 Appendix...
Страница 81: ...CS 7200 User Installation Guide 6M0940 _Ed04 77...
Страница 82: ...78 Chapter 8 Appendix...
Страница 84: ...80 Chapter 9 Contact Information...
Страница 86: ...82 Chapter 10 List of Consumables...