System Self Test
6K7304 | 2014-10-20
9–3
3. System performs Motors test.
A green check mark appears next to each tested function.
4. Expose one of the following cassettes: 14 x 17 in./10 x 12 in./24 x 30 cm. to a dose of
5 mR using x-ray settings: 60KVp, 1.5mAs, SID = 1.0 m. (40 in.). Then, insert the
cassette into the system.
The Image Readout Test starts.