Troubleshooting the CR System
8–4
6K7304 | 2014-10-20
Acceptance Test Procedure Failures
Symptom
Possible Cause
Solution
Need to perform System
Acceptance Test
•
Software was installed/rein-
stalled, but the Acceptance
Test Procedure was not
performed.
•
Service procedure,,such as
part replacement and/or
recalibration, was performed
by qualified technician.
Perform the Acceptance Test Proce-
dure.
Test failed due to “Wrong
angle detected”.
ATP phantom was not properly
aligned with the edges of the
cassette.
Properly align the edges of the ATP
phantom with the cassette edges.
Test failed due to “Wrong
exposure level detected”.
Measured Average Pixel Value
(APV) is not within the accepted
range of 5 +/- 1mR.
Readjust the X-ray settings according
to the displayed instructions:
Acceptance test has failed.
Please call Service.
The acceptance test has failed
for a reason other than wrong
exposure dose or phantom mis-
alignment.
•
Verify that the phosphor screen is
clean and undamaged.
•
Verify that the excessive light has
not penetrated the system:
Power supply and USB door is
properly closed.
No direct light source is pointed
toward the scanner.
Light level in the room is below 500
lux.
•
Call Service.
Содержание Carestream Vita Flex CR System
Страница 1: ...l Publication No 6K7304 2014 10 20 Version 2 0 CARESTREAM Vita Flex CR System User Guide...
Страница 6: ...iv 6K7304 2014 10 20...
Страница 8: ...About this Guide 1 2 6K7304 2014 10 20...
Страница 14: ...Site Preparation 2 6 6K7304 2014 10 20...
Страница 32: ...Install the CR System 4 16 6K7304 2014 10 20...
Страница 40: ...Use the CR System 5 8 6K7304 2014 10 20...
Страница 46: ...Use the CR System 5 14 6K7304 2014 10 20...
Страница 56: ...Maintain and Clean the System Cassette Screen Roller Cleaning 6 10 6K7304 2014 10 20...
Страница 82: ...System Self Test 9 6 6K7304 2014 10 20...
Страница 84: ...System Specifications 10 2 6K7304 2014 10 20...
Страница 86: ...Publications History II 6K7304 2014 10 20...