CR System Service Procedures
7–6
6K7304 | 2014-10-20
Note
Make sure that you release the clip before trying to remove the main power connector.
Install the Power Supply Cable
1. Connect the power supply cable and the USB cable to the CR System.
Note
You Must close the door to the power supply, if you do not close the door to the power supply
properly, there is a risk of light penetration resulting in loss of image quality
2. Install the cable compartment cover at the rear of the CR System. See
Install the Cable
Compartment Cover
3. Plug the power supply into the Mains wall socket
4. Switch on the CR System by pressing the On/Off button.
Important
Make sure that you install the door at the rear of the
CR System. If you do not install the
door at the rear of the CR System, light can penetrate the system and affect image quality.