Troubleshooting the CR System
6K7304 | 2014-10-20
8–9
Error code may vary
Scanner cannot operate.
General error
1. Short-press the On/Off button to
eject the cassette and reset the
scanner.
2. Verify that the scanner status is
Ready and the front panel LED is
green.
If the error persists, perform the follow-
ing steps:
1. Disconnect scanner from the power
supply.
2. Remove the Screen Transportation
Unit.
3. Visually inspect that there is no
screen jammed in the camera.
4. Insert the Screen Transportation Unit
into the system.
5. Connect the system to the power
supply.
6. Power up the scanner and verify that
the scanner status is Ready and the
front panel LED is green.
If the error persists, perform the follow-
ing steps:
1. Operate the System Self Test to diag-
nose the problem.
2. Call Service for further instructions.
Symptom
Possible Cause
Solution
Содержание Carestream Vita Flex CR System
Страница 1: ...l Publication No 6K7304 2014 10 20 Version 2 0 CARESTREAM Vita Flex CR System User Guide...
Страница 6: ...iv 6K7304 2014 10 20...
Страница 8: ...About this Guide 1 2 6K7304 2014 10 20...
Страница 14: ...Site Preparation 2 6 6K7304 2014 10 20...
Страница 32: ...Install the CR System 4 16 6K7304 2014 10 20...
Страница 40: ...Use the CR System 5 8 6K7304 2014 10 20...
Страница 46: ...Use the CR System 5 14 6K7304 2014 10 20...
Страница 56: ...Maintain and Clean the System Cassette Screen Roller Cleaning 6 10 6K7304 2014 10 20...
Страница 82: ...System Self Test 9 6 6K7304 2014 10 20...
Страница 84: ...System Specifications 10 2 6K7304 2014 10 20...
Страница 86: ...Publications History II 6K7304 2014 10 20...