Configured Options (Continued)
1. Select the
DRUG LIBRARY
option from the Configured Options
menu and press the
OK
softkey.
2. To enter a new drug in the library press the
NEW
softkey.
3. To enter the name of the new drug use the
f
keys to scroll
through the alphabet. To select a letter press the
NEXT
softkey. On
completion press the
OK
softkey.
4. Follow the flow chart, using the
f
keys to select values.
Use
OK
to enter selected values and move on to the next stage. The
BACK
softkey may be used at any time to return to the previous
screen of the drug library set-up procedure.
Dosing Conversion:
1.0 µg = 1000 ng
1.0 mg/h = 24.0 mg/24 h
1.0 mg/min = 60.0 mg/h
1.0 mg = 1000 µg
The availability of drug dose units is dependent on the software
version of the pump.
5. Review the drug set-up data on the display, then press the
OK
softkey to confirm.
6. To delete a drug from the drug library, select the drug using the
f
keys. Press
MODIFY
then select
YES
twice to delete the
drug from the list.
Note:
to set a FIXED PROTOCOL set the:-
Max Doserate = Default Doserate = Min Doserate;
Max Conc. = Default Conc. = Min Conc.
Note:
use of a protocol in which the Max Bolus is set OFF will prevent
bolus delivery.
Drug library