1000DF00329 Issue 4
15/36
Basic Features (Continued)
If the Auto Set Pressure Option is enabled then the pump AUTOMATICALLY adjusts the pressure occlusion limit.
l
After 15 minutes of infusion the pump AUTOMATICALLY adjusts the occlusion pressure limit to X mmHg (the AUTO OFFSET value, 15-
100mmHg), above the average infusion pressure since the start of the infusion.
Auto Set Pressure (If enabled)
Pressure Level with pressure set fitted
1. To check and adjust the pressure level press the
e
button. The display will change to show a 20 minute pressure trend graph displaying
the pressure alarm level and the current pressure level.
2. Press the
f
keys to increase or decrease the pressure alarm level. The new level will be indicated on the display.
3. The
AUTO
Pressure feature may be used when a stable pressure has been achieved over a short period of infusion. If
AUTO
Pressure
has been enabled the automatic pressure alarm level is calculated and set by pressing the
AUTO
softkey.
4. Press the
TREND
softkey to view the pressure trend of the previous 12 hours. The pressure trend can be viewed at 15 minute intervals
by using the
+/-
softkeys. The pressure trend graph displays the pressure at a given time.
5. Press the
OK
softkey to exit the pressure screen.
_
+
OK
20mmHg
07:45
Pressure Level without pressure set fitted
(not applicable when FULLY DEDICATED)
1. To check and adjust the pressure level press the
e
button. A bar graph will be displayed showing the pressure alarm level and the
current pressure level.
2. Press the
f
keys to increase or decrease the alarm level. The new level will be indicated on the display.
3. Press
OK
to exit the screen.
A
The interpretation of pressure readings and occlusion alarms are the responsibility of the clinician depending on the
specific application.
Rate Lock
If Rate Lock is enabled, when the infusion rate has been set and the infusion started (or following a bolus infusion) the rate lock prompt
will appear on the main display.
To select the rate lock function press the
YES
softkey. Press the
NO
softkey if the rate lock is not required.
When rate lock is enabled, the following are unavailable:
l
Changing the infusion rate / titration
l
Bolus / purge
l
Switching the pump off
l
VTBI over time infusions.
To disable the rate lock if selected:
1. Press the
d
button to access the options menu.
2. Select the
UNLOCK RATE
option using the
f
keys and press the
OK
softkey.
To enable the rate lock if not selected:
1. Press the
d
button to access the options menu.
2. Select
RATE LOCK
and press the
OK
softkey.