Configured Options
This menu comprises a list of options which are configurable by the user.
1. Turn the pump
OFF
.
2. Whilst holding down the
b
button turn the pump
ON
.
3. The main display will show
000
. Enter the access code for Configured Options using the
f
keys, pressing
NEXT
to move through
the digits. A full list of access codes can be found in the Technical Service Manual.
4. When the complete code shows on screen, press
OK
to enter. The Configured Options menu will be displayed.
1. Select
GENERAL OPTIONS
from the menu using the
f
keys and press the
OK
softkey.
2. Select the option you wish to enable/disable or adjust and press the
MODIFY
softkey.
3. When all the desired modifications have been carried out press the
QUIT
softkey.
4. Either select the next configuration option from the menu or turn the pump
OFF
, returning it to operation as required.
NURSE CALL FITTED
Enables Nurse Call (hardware option).
NURSE CALL INVERT
When enabled, the nurse call output is inverted.
RS232 SELECTED
Sets the pump's communications to use RS232 (hardware option).
NEOI WARNING
Sets the Near End Of Infusion warning time, as time left to End Of Infusion.
EOI POINT
Sets the End Of Infusion point.
KVO AT EOI
When enabled the pump will switch to running at the KVO rate when EOI is reached.
KVO RATE
Sets the Keep Vein Open (KVO) rate at which the pump will operate if KVO at EOI is enabled.
BACK OFF
When enabled the motor will reverse to relieve line pressure when an occlusion occurs.
AUTO SAVE
When disabled the infusion information is cleared on power up.
RATE LOCK
When enabled the rate can be locked to prevent unwanted changes of the set infusion rate.
QUIET MODE
When enabled the button beeps are muted.
AC FAIL
When enabled the AC Power Failure Alarm will sound if the AC power is disconnected.
RATE TITRATION
When enabled the rate can be changed whilst the pump is infusing.
PRESSURE DISPLAY
Enables / disables the Pressure Icon on the main display.
AUTO PRESSURE
Enables / disables the automatic pressure alarm level option.
AUTO SET PRESSURE
Automatically sets the line occlusion pressure to a specified amount above the current pressure.
AUTO OFFSET
Adjusts the automatic offset value used by auto pressure and auto set pressure. Adjustable from 15 to
100mmHg.
PRESSURE DEFAULT
Sets the default occlusion alarm level.
MAX PRESSURE
Sets the maximum pressure limit.
WEIGHT
Sets the default patient weight in kg.
CAP RATE
Sets the maximum value for infusion rate.
PURGE RATE
Sets the purge rate.
PURGE VOLUME LIMIT
Sets the maximum permissible purge volume.
PURGE SYRINGE
Prompt to purge syringe after confirmation.
BOLUS
Enables / disables the bolus feature.
DEFAULT BOLUS
Sets the default bolus rate.
CAP BOLUS RATE
Sets the maximum value for bolus rate.
BOLUS VOL LIMIT
Sets the maximum permissible bolus volume.
MANUAL BOLUS
Volume infused will be increased if plunger is manually moved in and syringe remains confirmed.
CALL BACK TIME
Adjusts the time for the pump to sound the call back alarm.
VTBI CLEAR RATE
Rate will be set to zero when VTBI has been set-up with stop as the end rate.
EVENT LOG DISPLAY
Enables / disables the event log.
BATTERY ICON
Enables / disables the Battery Icon on the main display.
AUDIO VOLUME
Sets the alarm volume of the pump at high, medium or low.
AUTO NIGHT MODE
Backlight dims between hours 21:00 and 06:00.
General Options