When using an overlay mattress system
(CuroCell S.A.M.® PRO):
1a. Open the cover and insert a hand, with the palm facing up, between the overlay mattress
and the underlying mattress. The hand is inserted beneath the user’s sacrum (center of
mattress).
When using a full
replacement mattress
(CuroCell S.A.M.® PRO CF16)
1b. Open the cover and insert a hand, with the palm facing up, between the overlay mattress
and the underlying mattress. The hand is inserted beneath the user’s sacrum (center of
mattress).
2. Ensure there is a gap between the user and the underlying mattress so that the user does
not ‘bottom out’.
3. If you can feel the user’s sacrum resting in the palm of your hand, the gap is too small.
3.4
Cardiopulmonary rescitation (CPR)
In case of an emergency where CPR (cardiopulmonary
resuscitation) is necessary:
Quickly pull the CPR strap on the air tubes between the pump
and the mattress. Detach the quick connectors on the side of
the pump, for even faster deflation.
3.5
Low pressure
If there is abnormally low air pressure, see section 8 for Troubleshooting.
3.6
Transport function
Detach both the tubes from the pump and connect these together; this will put the mattress in
“Transport mode”. We recommend for this function only be used for a short period of time.
Note!
•
Make sure that the mattress is filled, and that the system has adapted to
the user before performing a hand check.
•
How to perform a hand check depends on which mattress is used.
•
If the air pressure remains low, search for damage on the mattress
and the tubes. Repair when needed or contact the supplier.
Notera!
•
After use, make sure that
the CPR-conntection is
properly attached.