Operation

 

 

3.1

 

Operation (how to get started) 

1. Start the pump by pressing “1” (on), on the side of the pump. 

 

2. The pump starts filling the mattress with air. The green light with the symbol  

will 

flash during the inflation phase; it takes about 15 minutes for the mattress to fill with air. 

 

3. When the 

symbol shows a steady green light, the mattress is ready to be used. 

 

 

 

 

3.2

 

Comfort settings 

 

The air pressure in the mattress can be adjusted according to the user’s comfort preference. 

Always consult the prescriber or other authorised personnel for the appropriate setting. 

 

Appropriate air pressure is selected using the knob (0–250 kg). The numbered scale guides 

your selection. For sitting positions, we recommend that the pressure be raised by +25 kg. 

Check using the hand check (see section 3.3). 

 

 

 

 

3.3

 

Hand check (function control) 

 

Hand check is performed to ensure that the mattress 

system works properly and to ensure that the weight 

setting is correct. Hand check should be performed 

regularly; for S.A.M Pro, we recommend once per work 

shift or every 8 hours, as well as after installation of the 

system. 

 

 

For manual systems hand check should also be 

performed when moving from lying to seating position in 

bed, in case of positioning changes and when changing 

the comfort setting (and/or the weight setting).

  

 

Note! 

 

• Make the bed with bedsheets.  

 

•It is recommended to not use the product during x-ray.  

 

•If the product has been stored in a cold space, wait some time using it. 

 

Note! 

 

• When the head end of the bed is raised, the pressure should be increased by 25 kg. 

 

• When only part of the mattress is under load, e.g. for amputee users, the pressure 

should be increased (by 1–2 steps). Then perform a hand check (see section 3.3). 

 

 

Содержание CuroCell S.A.M. PRO

Страница 1: ...EN Instructions for use 2022 06 23 CuroCell S A M PRO Constant low pressure reactive air mattress system Instructions for use item number 95 001410 EN0000 ...

Страница 2: ...heck function control 6 3 4 Cardiopulmonary rescitation CPR 7 3 5 Low pressure 7 3 6 Transport function 7 4 Product description 8 4 1 Pump 8 4 2 Control panel 9 4 3 Mattress 10 5 Reuse and cleaning 10 5 1 Cleaning pump 10 5 2 Mattress cover 10 5 3 Inner cover 11 6 Storage 12 7 Maintenance 12 7 1 Air filter and fuse replacement 12 8 Troubleshooting 13 9 Technical specification 14 9 1 Standards 14 9...

Страница 3: ... defective the pump casing is damaged or it is not functioning properly Contact an authorised service technician for examination and repair 6 Regularly check product functionality through a hand check see section 3 4 Also check that the comfort setting is correct as indicated in the table on the pump 7 Continual monitoring is required when the product is used for individuals needing special superv...

Страница 4: ...n individualised plan of care The mattress system is available in three versions CuroCell S A M PRO and CuroCell S A M PRO CF10 are top mattresses and are placed on an underlying mattress CuroCell S A M PRO CF16 is a replacement mattress placed directly on the bed base The recommended user weight is up to 250 kg 1 6 Contraindication It is necessary for the prescriber of the mattress to make an ind...

Страница 5: ...mp by the foot end of the bed or place it on a flat sturdy surface Place the power cable so there is no risk of stumbling over it run it over with the bed wheels or getting it jammed when raising or lowering the bed 3 Connect the air tubes coming from the mattress to the side of the pump The pump and hoses are equipped with quick connectors an audible click will be heard when these have been conne...

Страница 6: ...k using the hand check see section 3 3 3 3 Hand check function control Hand check is performed to ensure that the mattress system works properly and to ensure that the weight setting is correct Hand check should be performed regularly for S A M Pro we recommend once per work shift or every 8 hours as well as after installation of the system For manual systems hand check should also be performed wh...

Страница 7: ...ation CPR In case of an emergency where CPR cardiopulmonary resuscitation is necessary Quickly pull the CPR strap on the air tubes between the pump and the mattress Detach the quick connectors on the side of the pump for even faster deflation 3 5 Low pressure If there is abnormally low air pressure see section 8 for Troubleshooting 3 6 Transport function Detach both the tubes from the pump and con...

Страница 8: ...setting knob 2 Control panel 3 Casing front 4 Tube connector 5 Power cable connector 6 Power switch Used to turn the pump on or off A green light indicates when the pump is on 7 Casing back A Hangers B Fuses 2 x T1A 250 V C Label D Air filter ...

Страница 9: ...vides fixed support This function should be used when nursing the user shifting the user s position or moving the user in or out of bed This function automatically reverts to the previous setting after about 30 minutes Information signal If the mattress pressure is too low an audio visual alarm is triggered Turning off the information signal Turning off the information signal Turn off the audio al...

Страница 10: ...en peroxide max 1 5 If using a different agent choose one that does not affect the pump 5 2 Mattress cover Covers can be wiped off with a mild cleaning solu tion for this purpose such as alcohol with or with out tenside surfactants Isopropanol 70 oxidizing solutions chlorine max 1 max 10 for cover Stone or hydrogen peroxide max 1 5 Excrement and blood stains must be removed as soon as possible usi...

Страница 11: ...olutions chlorine max 1 or hydrogen peroxide max 1 5 Let it dry 5 Wipe off the work surfaces using a suitable detergent and or disinfectant 6 Put the mattress together If cells for any reason have become loose from the tubes these must be replaced according to the drawing in section 4 3 Foam Core S A M Pro CF16 Clean the affected area with a mild detergent such as washing up liquid and water or wi...

Страница 12: ...nd power supply for damage and scratches 2 Check the mattress cover to make sure it is undamaged and that covers and cells are correctly assembled 3 Check that air is coming from the outlets on the side of the pump 4 Check that no tubes or connectors are damaged or pinched Contact the manufacturer or your local distributor for spare parts 7 1 Air filter and fuse replacement Air filter A 1 Loosen t...

Страница 13: ...hat the air filter is not blocked The user is bottoming out The comfort settings might be incorrectly set Adjust the pressure and wait a few minutes Check again using the hand check see section 3 3 Check that the CPR valve is closed Mattress not stationary Check that the mattress is secured to the underlying mattress with the elastic straps or if the mattress has an integrated foam core CuroCell S...

Страница 14: ...er size care and wash instructions EAN bar code as well as space for own labelling The product also displays which side is up down and head foot PUMP SPECIFICATION Model CuroCell S A M PRO Voltage 110 120 V 60 Hz 220 240 V 50Hz Current Max 10 W Fuse type T1A 250V Dimensions LxWxH 26 x 8 x 13 cm Weight 1 5 kg Sound level max ISO 3746 2010 39 dBA EN ISO 11201 2010 22 dBA Ambient environment Temperat...

Страница 15: ...ipment double insulated Indicated on the power supply Symbols for user information Recommended user weight XXXX XX XX Year month day Counteracts shear force User information category Placed directly on the bed base User information category Foot placement Can be rotated Heel function Can be turned The mattress should be used with the user lying lengthways Place on top of existing mattress Warning ...

Страница 16: ...eaning and recycling Machine wash at 95 C Do not iron Do not machine wash Chlorine Wipe clean Tumble dry Do not dry clean Do not tumble dry Drip dry Do not dispose of with household waste Recycling Machine wash at 70 C ...

Страница 17: ...ineration in waste incineration facilities Pump A used CuroCell pump must not be disassembled Instead it should be taken to a recycling centre The product is sorted as electronic waste Mattress A used CuroCell mattress should be taken to a recycling centre The product is sorted as combustible waste For more information please contact Care of Sweden or your local distributor 10 3 Returns Contact Ca...

Страница 18: ...18 Your own notes ...

Страница 19: ...19 ...

Страница 20: ...se Internet www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Visit address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns väg 4A SE 514 33 Tranemo SWEDEN Distributed by Care of Sweden is a trademark of Care of Sweden AB Care of Sweden AB 2022 ...

Отзывы: