11
5.3
Inner cover
Clean the affected area with a mild cleaning solution for this purpose such as: alcohol
with or without tenside surfactants, Isopropanol 70 %, oxidizing solutions, chlorine max
1 % or hydrogen peroxide max 1,5 %.
5.4
Mattress
1. Disconnect the tube connector from the pump and remove the air from the mattress.
2. Remove the covers.
3. Move the mattress to a clean area that is suitable for cleaning.
4. Wipe off the cells, all tubes and the CPR module with a mild cleaning solution for this
purpose such as: alcohol with or without tenside surfactants, Isopropanol 70 %, oxidizing
solutions, chlorine max 1 % or hydrogen peroxide max 1,5 %. Let it dry.
5. Wipe off the work surfaces using a suitable detergent and/ or disinfectant.
6. Put the mattress together. If cells for any reason have become loose from the tubes, these
must be replaced according to the drawing in section 4.3.
Foam Core (S.A.M Pro CF16)
Clean the affected area with a mild detergent (such as washing-up liquid) and water or with an
alcohol-based disinfectant (cleaner intended for this purpose. Gently squeeze out any water.
Note!
• Check the whole product for damage each time it is cleaned. If damaged, it must be
replaced or repaired.
•Do not wring or roll the foam core to extract the water. Let it dry in a warm, ventilated
area (not in direct sunlight). The foam core must be completely dry before it is used again.