11
5 Cleaning and reconditioning
It is important to follow the instructions below before using the product for another user.
Consult the hygiene responsible or the supplier for help and instructions when unsure.
5.1 Cleaning
PUMP
Wipe of the pump with a damp cloth and a mild cleaning solution for this purpose
such as: alcohol with or without tensides, Isopropanol 70 %, oxidating solutions or
chlorine max 1 %. If using a different agent, choose one that does not affect the plastic
of the pump housing.
MATTRESS COVER
Covers can be wiped off with a mild cleaning solution for this purpose such as: alcohol
with or without tensides, Isopropanol 70 %, oxidating solutions or chlorine max 1 %.
Excrements and blood spots must be removed immediately with cold water. Carefully
follow local instructions and usage instructions for the cleaner.
Multi-layer covers should be divided before washing. W
ash with like colors.
Chlorine and phenol-based cleaners could have a negative impact on the PU surface
and should be avoided. If chlorine is used, we recommend a mixture of max. 1%.
INNER COVER AND DISTANCE FABRIC
(accessories)
Clean the affected area with a mild cleaning solution for this purpose such as:
alcohol with or
without tensides, Isopropanol 70 %, oxidating solutions or chlorine max 1 %.
WARNING!
Do not open the pump housing – risk of electric shock. Service and maintenance must be
performed by authorised technicians. Do not get the pump wet or submerge it in any
liquid.