7
2.3 Pump functions – control panel
/
Button for switching between static and alternating function.
= M
ode 1 = The pressure change between cells, in a cycle of ten minutes.
We recommend that this mode is used by default.
=
Mode 0 = Same pressure in all cells. In static mode the alternatating
cycle is discontinued and the pressure is equalized over the cells.
This
function makes all user movements, as well as nursing and care of the
user, easier.
Power off/on (1/0).
/
Suitable air pressure is set with the comfort dial (1-8). Guidelines based
on user weight is given in the table on the pump.
NOTE!
In all uses
where the head end of the bed is raised,
the pressure must be increased
at least one (1) step.
Lights when the mattress is ready for use.
Alarms for low pressure. In order to solve it see chapter 8,
troubleshooting.