4.4 Hand check (function control) 

 

The hand check is used to ensure that the mattress is 

working properly. This must be performed regularly; we 

recommend once per shift, or every eight hours. 

 

Place a flat hand under the air cells directly underneath 

the user’s pelvic region and make sure that the user 

does not lay directly on the underlying mattress. There 

should be about 1-2 fingers margin.  

 

If you can feel the user´s body resting on your hand you need to increase the comfort settings 

to next highest step. Wait 10 minutes and perform a new hand check. 

 

4.5 CPR 

 

In case of an emergency where CPR (cardiopulmonary 

resuscitation) is necessary: Rapidly pull the CPR band. 

Remove the tube connections on the side of the pump, 

for even faster deflation.  

 

NOTE!

 Make sure that the CPR connector is properly 

connected after use. 

 

4.6 Low pressure  

 

At an abnormally low air pressure in the mattress,

 

the symbol for "Low pressure" lights up        .

 

         

. The lamp goes out when the pressure returns to normal. 

 

Please refer to chapter 8 for troubleshooting. 

 

 

 

 

 

 

 

WARNING! 

If the air pressure remains low, check the mattress and tubes for damage. Repair as 

necessary or contact Care of Sweden or one of its authorised repair technicians. 

 

Содержание CuroCell Nova

Страница 1: ...User manual 2021 03 04 CuroCell Nova Alternating pressure mattress Item number user manual 95 001014...

Страница 2: ...on 4 Product description 6 Assembly Installation 8 Operation 9 Cleaning and reconditioning 11 Storage 12 Maintenance 13 Troubleshooting 14 Technical description 15 Spare parts 17 Options and accessori...

Страница 3: ...t in a manner that prevents it from becoming trapped or damaged NOTE Be particularly aware of trapping damage when using bed rails 5 Never use the product if the pump power cord or plug is defective t...

Страница 4: ...to our products by reporting them immediately to us here at Care of Sweden The system has been tested and approved according to the following standards IEC 600601 1 IEC 60601 1 2 SS EN ISO 12182 SS E...

Страница 5: ...Customer Services see contact information on the final page NOTE For certain patients such as amputees are the recommended minimum weight not valid These users may however have other requirements as...

Страница 6: ...n 2 1 Pump and mattress 1 Mattress 2 CPR rapid deflation 3 Tubes 4 Pump connector 5 Pump Air cells 2 2 Pump 1 Connection tubes 3 Power cord 5 Suspension hook 2 On Off 4 Fuses 2 6 Filter 2 4 1 3 4 3 1...

Страница 7: ...ed and the pressure is equalized over the cells This function makes all user movements as well as nursing and care of the user easier Power off on 1 0 Suitable air pressure is set with the comfort dia...

Страница 8: ...2 Hang the pump at the foot end of the bed or place it on a flat sturdy surface Route the power cord so that there is no risk of it causing someone to trip being run over by the bed wheels or getting...

Страница 9: ...programs see section 2 3 for details We recommend alternating mode by default 4 3 Comfort setting The air pressure in the mattress can be adjusted according to the users desire for comfort Always cons...

Страница 10: ...highest step Wait 10 minutes and perform a new hand check 4 5 CPR In case of an emergency where CPR cardiopulmonary resuscitation is necessary Rapidly pull the CPR band Remove the tube connections on...

Страница 11: ...Isopropanol 70 oxidating solutions or chlorine max 1 Excrements and blood spots must be removed immediately with cold water Carefully follow local instructions and usage instructions for the cleaner...

Страница 12: ...g solutions or chlorine max 1 5 Wipe off and disinfect all the work surfaces with an alcohol based cleaning agent intended for this purpose 6 Put the mattress together by putting back the cell package...

Страница 13: ...s 7 2 Changing fuses 1 Pull the plug out of the electrical socket before replacing fuses 2 Remove the cover for the fuses with a screwdriver Pull out the fuse holder 3 Fit a new fuse of the same type...

Страница 14: ...check see section 4 4 Mattress moves around Check that the air mattress is secured to the underlying mattress with the elastic straps Some cells have less air This is normal since the air flow alterna...

Страница 15: ...ospheric pressure 700hPa 1013 25hPa Classification Class II Type BF IP2X MATTRESS SPECIFICATION Model CuroCell Nova Dimension LxWxH 85x200x13 cm Weight 4 kg NOTE Care of Sweden retains the right to mo...

Страница 16: ...CE marked acc to Medical Device Regulation MDR 2017 745 Machine wash at 95 C Tumble dry Place on top of existing mattress Do not iron Recommended user weight User information category Wipe off Year M...

Страница 17: ...ries 11 1 Optionals PRIMA 6 An elevating pad that is placed underneath a mattress overlay for example CuroCell Nova air mattress 11 2 Accessories Art no Description CC 9091 Sturdy transport bag 11 3 O...

Страница 18: ...e of its authorised technicians Only use spare parts that have been approved by Care of Sweden Use of non approved spare parts invalidates the warranty A warranty claim must be submitted to Care of Sw...

Страница 19: ...19 14 Own notes...

Страница 20: ...ternet www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Visit address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns v g 4A SE 514 33 Tranemo SWEDEN Sup...

Отзывы: