7. Maintenance
7.1 General
1. Check the power cord and plug for damage and scratches.
2. Check that the mattress cover is intact and that the cover and air cells are correctly
assembled.
3. Check that the air flow alternates between each outlet on the side of the pump.
4. Check for damaged or pinched tubes and connectors.
Contact the manufacturer or your local dealer for spare parts.
7.2 Changing fuses
1. Pull the plug
out of the electrical socket
before replacing
fuses.
2. Remove the cover for the fuses with a screwdriver. Pull out
the fuse holder.
3. Fit a new fuse of the same type (T1A/250 V) and screw
the protective cover back in place. There are two fuses.
7.3 Changing the air filter
1. Loosen the small round cover plate at the back of the pump.
2. Lean the pump back and the filter slides out of its holder.
3. Place the new filter in the holder and put back the cover
plate.
The filter should be cleaned regularly if the pump is used in a
very dirty environment.