8
3.
The light on the right is lit – The pump will automatically adjust its settings within five
minutes.
Once the settings have been automatically adjusted, the system will revert to the previously
used programme. However, when started
it will switch to Alternating mode by default.
To stop the automatic setting adjustments, press the button.
2. Comfort setting
The pressure can be increased or decreased one level depending on the user’s comfort
requirements.
This function cannot be used when the Sit function programme is activated.
3. Programmes
Four programmes are available. Repeatedly pressing the button switches between the
different programmes.
1.
Alternating mode. Here, the pressure alternates between the cells.
2.
CLP mode. Here, the pressure will be same in all cells. This is also called “Constant low
pressure”.
3.
Sit function. Gives the user more support in a sitting position. This function can be used in
both Alternating and CLP mode.
4. Maximun pressure (caring mode)
With this function, the entire mattress is inflated and provides firm support. This function
automatically reverts to the previous setting after about 24 minutes.
NOTE:
The automatic
settings are overridden when using this programme.
5. Panel lock
Press the Panel lock button to lock or unlock the panel. The green light is lit when the panel is
locked. The panel locks automatically after five minutes of inactivity. This is to prevent the
settings being changed inadvertently.
6. Power off/on
Turns the power to the control panel on or off.
NOTE:
Check that the power cable is properly
connected and that the power switch on the side of the pump is in position 1 (on).
7. Notification
In the event of problems with notification, see section 8 for troubleshooting.
A. Power supply notification
If the pump loses power an audible and visual notification
is triggered. Turn off the notification
by pressing the “Notification off"
button.
B. Low pressure notification
If the mattress pressure is too low an audible and visual notification is triggered. Turn off the
notification by pressing the “Notification off” button. It is reactivated after about five minutes
if the pressure has not returned to normal.
Alternating
CLP
Alter Sit function
CLP + Sit function