15
6. Storage
It is advisable to store the mattress and pump in the product bag (accessory). Handle the
stored product with care. Do not place any heavy objects on top of it.
7. Maintenance
7.1 General
1. Check the power cable and plug for damage and scratches.
2. Check the mattress protection to make sure it is undamaged and that covers and cells are
correctly assembled.
3. Check that the air flow alternates between the different outputs on the side of the pump.
4. Check that no tubes or connectors are damaged or pinched. Contact Care of Sweden or your
local distributor if any spare parts are required.
7.2 Changing fuses
1. Remove the plug from the wall socket when replacing
fuses.
2. Use a screwdriver to remove the fuse cover. Pull out
the fuse holder.
3. Insert a new fuse of the same type (T1A/250V) and
screw the protective cover back into place. There are
two fuses.
7.3 Changing the air filter
1. Loosen the small, round protective plate on the back of
the pump.
2. Remove the filter from the holder.
3. Insert the new filter in the holder and replace
the protective plate.
If the pump is used in a very dirty environment, the filter
should be checked regularly.