Care of Sweden CuroCell AUTO4 Скачать руководство пользователя страница 6

6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(1) 

European Pressure Ulcer Advisory Panel, National Pressure Injury Advisory Panel and Pan Pacific Pressure Injury 

Alliance. Prevention and Treatment of Pressure Ulcers/Injuries: Clinical Practice Guideline. The International Guideline. 

Emily Haesler (Ed). EPUAP/NPIAP/PPPIA: 2019

 

 

1.6 Other information 

 

Action plan for pressure ulcers/pressure injuries 

For the best possible results when using this product, we recommend a structured and 

planned approach. Examples that can be applied to the use of our products can be found in 

the “Action plan for pressure ulcers/pressure injuries” guidelines. 

 

The Mattress Guide 

This guide provides guidance to personnel and prescribers when choosing a mattress from 

Care of Sweden. CuroCell® AUTO4/AUTO4 CX13 is a mattress in function group C. This 

information can be obtained from our website, www.careofsweden.com, or ordered from 

Customer Service (see contact information on the last page). 

 

 

2. Product description 

 

2.1 Mattress 

 

1. Mattress 

2. CPR (Rapid deflation) 

3. Pump connector 

4. Pump 

 

 

 

 

 

NOTE: 

 

For certain users, such as amputees, the recommended height measurement may not 

be met. Users in these groups may, however, require other settings, as the entire 

surface is not under load. To check functionality, see sections 4.2, 4.3 and 4.4. 

 

 

When used together with positioning pillows, read the pillow instructions for use 

carefully for correct positioning of the user. 

 

 

Whenever this product is used with an evacuation sheet, it is the responsibility of the 

attending personnel to ensure that evacuation can be performed safely. 

 

 

Be careful when using bed rails or other protection on the bed so that the mattress is 

not pinched or damaged. 

 

 

The mattress must not be lifted with a person lying on it or used for any other type of 

transport other than necessary evacuation using an evacuation sheet.

 

 

 

The mattress is not suitable for use during X-ray examinations.

 

This is because of the 

risk of image blurring and artefacts that may lead to

 

diagnostic errors.

 

 

 

 

 

Содержание CuroCell AUTO4

Страница 1: ...Instructions for use 2020 12 08 CuroCell AUTO4 CuroCell AUTO4 CX13 Automatic active air mattress systems Instructions for use item number 95 001616...

Страница 2: ...duct description 6 3 Assembly Installation 9 4 Operation 10 5 Reuse cleaning and reconditioning 13 6 Storage 15 7 Maintenance 15 8 Troubleshooting 16 9 Technical description 17 10 Spare parts 19 11 Op...

Страница 3: ...functionality by performing a hand check see section 4 4 4 When the product is used for individuals needing special supervision such as children continuous monitoring is required 5 The mattress is pro...

Страница 4: ...least 1 meter from the control unit 15 Never use the product if the power cable plug of the control unit or power supply housing is defective if the control unit housing is damaged or if it is not fun...

Страница 5: ...authorities regarding medical devices the manufacturer is required to report all accidents or incidents involving the products We would be very grateful for all information involving accidents or inc...

Страница 6: ...information on the last page 2 Product description 2 1 Mattress 1 Mattress 2 CPR Rapid deflation 3 Pump connector 4 Pump NOTE For certain users such as amputees the recommended height measurement may...

Страница 7: ...automatically sets the pressure based on the user s weight posture and position There are three ways to use this function 1 The system adjusts the settings automatically when the system has been start...

Страница 8: ...is inflated and provides firm support This function automatically reverts to the previous setting after about 24 minutes NOTE The automatic settings are overridden when using this programme 5 Panel l...

Страница 9: ...y to the moving parts of the bed Check the cells and press studs to ensure they are correctly assembled 2 Hang the pump on the foot end of the bed or place the pump on a firm level surface Place the p...

Страница 10: ...4 2 Programme selection Four different programmes are available See section 2 3 for detailed information We recommend Alternating mode as the default setting 4 3 Comfort setting fine tuning The air p...

Страница 11: ...pump control panel If the problem is not resolved within about five minutes the notification will be reactivated The pump is also equipped with a power supply notification If the pump loses power an a...

Страница 12: ...for the user s heels This is achieved by disconnecting shutting down the cell in the mattress directly under the user s heel s It will no longer be inflated and will create a space between the cells...

Страница 13: ...oes not affect the pump MATTRESS COVER Covers can be wiped off with a mild cleaning solu tion for this purpose such as alcohol with or with out tenside surfactants Isopropanol 70 oxidizing solutions c...

Страница 14: ...ergent and or disinfectant 6 Put the mattress together If cells for any reason have become loose from the tubes these must be replaced according to the drawing in section 2 1 FOAM CORE Clean the affec...

Страница 15: ...ubes or connectors are damaged or pinched Contact Care of Sweden or your local distributor if any spare parts are required 7 2 Changing fuses 1 Remove the plug from the wall socket when replacing fuse...

Страница 16: ...sections 4 3 and 4 4 for more Mattress not stationary Check that the mattress is secured with the elasticated straps to the underlying mattress or if the mattress has an integrated foam cushion secure...

Страница 17: ...NB Care of Sweden reserves the right to modify the product specification at any time 9 2 Standards The system is tested and approved according to the following European standards where applicable req...

Страница 18: ...CE marked in accordance with Medical Device Regulation MDR 2017 745 Machine wash at 95 C Placed directly on the bed base Tumble dry Do not iron Recommended user weight User information category Wipe c...

Страница 19: ...PU 80x13 cm CC 901002 K Air cell PU 85x13 cm CC 901003 K Air cell PU 90x13 cm CC 901004 K Air cell PU 105x13 cm CC 901005 K Air cell PU 120x13 cm CC 8217 Fuse 2 pack CC 400903 Pump for mattress width...

Страница 20: ...4 105 120 cm AUTO4 CX13 Sheet holder each Only AUTO4 Item no CC 9075 Item no Description CC 900220 Heel function with valve 4 cells CC 900222 Heel function with valve 3 cells 11 3 Other accessories To...

Страница 21: ...regarded as warranty faults are rectified free of charge A warranty claim must be submitted to Care of Sweden before the product is returned 12 3 Recommended life time of the product This product s ex...

Страница 22: ...22 14 Notes...

Страница 23: ...23...

Страница 24: ...se Internet www careofsweden com Postal address Care of Sweden AB Box 146 SE 514 23 Tranemo SWEDEN Visiting address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Deliveries address Byns v g 4A SE 514 33 T...

Отзывы: