Care of Sweden CuroCell AUTO4 Скачать руководство пользователя страница 5

5

 

 

1. Introduction

 

 

CuroCell® air mattresses work according to the alternating pressure principle, a technique that 

is thoroughly tried and tested and has been used for a long time. Alternating pressure means 

that the pressure is alternated regularly between the cells in order to simulate the body’s own 

movements. Periods are created in which the pressure is lower or completely absent. This 

provides good conditions for the skin to recover and to prevent injury. 

               Always read the instructions for use prior to use. 

 
For further questions regarding installation, use or maintenance please contact Care of 

Sweden or your local distributor. 

1.1 General information 

This system is a medical device with CE marking in accordance with Medical Device Regulation 

MDR (2017/745). 
The system has been fully or partially designed and verified in accordance with the standards 

listed in section 9.2. Standards. 
According to statutory regulations made by the authorities regarding medical devices, the 

manufacturer is required to report all accidents or incidents involving the products. We would 

be very grateful for all information involving accidents or incidents relating to our products, by 

reporting them immediately to us, at Care of Sweden. 

1.2 Intended use 

The system consists of a pump and a mattress and are used for prevention and as an aid in the 

treatment of pressure ulcers/pressure injuries (PU/PI).   

1.3 Indications 

For users with pressure ulcers/pressure injuries (PU/PI) up to and including category 4 and 

unclassified PU/PI (unstageable and suspected deep tissue injury¹). As well as for users in risk 

of PU/PI.  

1.4 Contraindications 

There are no known contraindications. 

1.5 User application 

The systems can be used in all kinds of health care environments.

 

It is available in two 

versions. CuroCell

®

 AUTO4 CX13 is an overlay mattress and is placed on an underlying 

mattress. CuroCell

®

 AUTO4 is a replacement mattress placed directly on the bed base. The 

mattress 

is intended to be used for people with a recommended minimum height of 120 cm 

and a recommended user weight of up to 180 kg. 
 

 

Содержание CuroCell AUTO4

Страница 1: ...Instructions for use 2020 12 08 CuroCell AUTO4 CuroCell AUTO4 CX13 Automatic active air mattress systems Instructions for use item number 95 001616...

Страница 2: ...duct description 6 3 Assembly Installation 9 4 Operation 10 5 Reuse cleaning and reconditioning 13 6 Storage 15 7 Maintenance 15 8 Troubleshooting 16 9 Technical description 17 10 Spare parts 19 11 Op...

Страница 3: ...functionality by performing a hand check see section 4 4 4 When the product is used for individuals needing special supervision such as children continuous monitoring is required 5 The mattress is pro...

Страница 4: ...least 1 meter from the control unit 15 Never use the product if the power cable plug of the control unit or power supply housing is defective if the control unit housing is damaged or if it is not fun...

Страница 5: ...authorities regarding medical devices the manufacturer is required to report all accidents or incidents involving the products We would be very grateful for all information involving accidents or inc...

Страница 6: ...information on the last page 2 Product description 2 1 Mattress 1 Mattress 2 CPR Rapid deflation 3 Pump connector 4 Pump NOTE For certain users such as amputees the recommended height measurement may...

Страница 7: ...automatically sets the pressure based on the user s weight posture and position There are three ways to use this function 1 The system adjusts the settings automatically when the system has been start...

Страница 8: ...is inflated and provides firm support This function automatically reverts to the previous setting after about 24 minutes NOTE The automatic settings are overridden when using this programme 5 Panel l...

Страница 9: ...y to the moving parts of the bed Check the cells and press studs to ensure they are correctly assembled 2 Hang the pump on the foot end of the bed or place the pump on a firm level surface Place the p...

Страница 10: ...4 2 Programme selection Four different programmes are available See section 2 3 for detailed information We recommend Alternating mode as the default setting 4 3 Comfort setting fine tuning The air p...

Страница 11: ...pump control panel If the problem is not resolved within about five minutes the notification will be reactivated The pump is also equipped with a power supply notification If the pump loses power an a...

Страница 12: ...for the user s heels This is achieved by disconnecting shutting down the cell in the mattress directly under the user s heel s It will no longer be inflated and will create a space between the cells...

Страница 13: ...oes not affect the pump MATTRESS COVER Covers can be wiped off with a mild cleaning solu tion for this purpose such as alcohol with or with out tenside surfactants Isopropanol 70 oxidizing solutions c...

Страница 14: ...ergent and or disinfectant 6 Put the mattress together If cells for any reason have become loose from the tubes these must be replaced according to the drawing in section 2 1 FOAM CORE Clean the affec...

Страница 15: ...ubes or connectors are damaged or pinched Contact Care of Sweden or your local distributor if any spare parts are required 7 2 Changing fuses 1 Remove the plug from the wall socket when replacing fuse...

Страница 16: ...sections 4 3 and 4 4 for more Mattress not stationary Check that the mattress is secured with the elasticated straps to the underlying mattress or if the mattress has an integrated foam cushion secure...

Страница 17: ...NB Care of Sweden reserves the right to modify the product specification at any time 9 2 Standards The system is tested and approved according to the following European standards where applicable req...

Страница 18: ...CE marked in accordance with Medical Device Regulation MDR 2017 745 Machine wash at 95 C Placed directly on the bed base Tumble dry Do not iron Recommended user weight User information category Wipe c...

Страница 19: ...PU 80x13 cm CC 901002 K Air cell PU 85x13 cm CC 901003 K Air cell PU 90x13 cm CC 901004 K Air cell PU 105x13 cm CC 901005 K Air cell PU 120x13 cm CC 8217 Fuse 2 pack CC 400903 Pump for mattress width...

Страница 20: ...4 105 120 cm AUTO4 CX13 Sheet holder each Only AUTO4 Item no CC 9075 Item no Description CC 900220 Heel function with valve 4 cells CC 900222 Heel function with valve 3 cells 11 3 Other accessories To...

Страница 21: ...regarded as warranty faults are rectified free of charge A warranty claim must be submitted to Care of Sweden before the product is returned 12 3 Recommended life time of the product This product s ex...

Страница 22: ...22 14 Notes...

Страница 23: ...23...

Страница 24: ...se Internet www careofsweden com Postal address Care of Sweden AB Box 146 SE 514 23 Tranemo SWEDEN Visiting address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Deliveries address Byns v g 4A SE 514 33 T...