5.2 Reconditioning
PUMP
See section 5.1 for instructions.
MATTRESS
1.
Disconnect the tube connector from the pump and remove the air from the
mattress.
2.
Remove the covers. For more information of cleaning the covers, see section 5.1.
3.
Move the mattress to a clean area that is suitable for cleaning.
4.
Wipe off the cells, all tubes and the CPR module with a mild cleaning solution for this
purpose such as: alcohol with or without tenside surfactants, Isopropanol 70 %, oxidizing
solutions, chlorine max 1 % or hydrogen peroxide max 1,5 %. Let it dry.
5.
Wipe off the work surfaces using a suitable detergent and/ or disinfectant.
6.
Put the mattress together. If cells for any reason have become loose from the tubes,
these must be replaced according to the drawing in section 2.1.
FOAM CORE
Clean the affected area with a mild detergent (such as washing-up liquid) and water or with an
alcohol-based disinfectant /cleaner intended for this purpose.
Gently squeeze out any water.
6. Storage
NOTE:
Check the hygiene cover, cells and hoses each time the product is
cleaned.
If damaged, it must be replaced or repaired. Also check the pump, tube connectors
and power cable during cleaning. Damaged parts must be replaced or repaired.
Do not wring or roll the foam core to extract the water. Let it dry in a warm, ventilated area
(not in direct sunlight). The foam core must be completely dry before it is used again.