Reference table for the connection to the patient
IEC lead
AAMI lead
Position
C1
Red
V1
Red
Fourth intercostal space to the right of the
sternum.
C2
Yellow
V2
Yellow
Fourth intercostal space to the left
of the sternum.
C3
Green
V3
Green
Midway position between electrodes V2/C2
and
V4/C4.
C 4
Brown
V4
Blue
Fifth intercostal space at midclavicular line.
C5
Black
V5
Orange
Between electrodes V4 and V6.
C6
Violet
V6
Violet
Level with electrode V4 at left midaxillary line.
L
Yellow
LA
Black
On the deltoid muscle, the forearm and wrist.
R
Red
RA
White
On the deltoid muscle, the forearm and wrist.
F
Green
LL
Red
On the thigh or the ankle.
N
Black
RL
Green
On the thigh or the ankle.
7.3.
ECG acquisition
Use the LED indicator to check the proper connection between the HD+ and the receiver.
The LED flashing indicator signals that the HD+ is attempting to connect to the receiver but is not yet
connected. The steady LED indicator signals that the HD+ is properly connected to the receiver.
Follow the instructions contained in Chapter 6 to prepare the device. Specifically:
1.
Insert two AAA batteries into the dedicated compartment. If the battery is too low, the HD+ may not
power on. Insert two new batteries into the device to continue the operation.
C1
C4
C2
C3
C5
C6
L
V1
V2
V3
V4
V5
V6
LA
R
N
RR
AA
RL
F
LL
Содержание HD+
Страница 1: ...1936 Rev 14 23 04 2019 HD User manual...
Страница 6: ...HD 1 GENERAL INFORMATION 2...
Страница 12: ...HD 2 SAFETY INFORMATION 8...
Страница 18: ...HD 4 SYMBOLS AND LABEL 14...
Страница 26: ...HD 5 INTRODUCTION 22...
Страница 28: ...HD 6 PREPARATION FOR USE 24...
Страница 40: ...HD 9 TECHNICAL SPECIFICATIONS 36...
Страница 42: ...HD 10 WARRANTY 38...
Страница 44: ...HD 40...
Страница 45: ...HD 41...
Страница 46: ...HD 42...
Страница 47: ...HD 43...