Air
Life
®
nCPAP Driver Service Manual
2008, Cardinal Health
Page 37 of 81
These procedures should be followed in sequence to provide easiest disassembly when access to
driver interior is desired.
12.1
Remove the Top Case from the Bottom Case
1.
Turn the unit upside down on a clean, dry, level surface.
2.
Use a 9/64” Allen wrench to loosen and remove the four (4) #8 screws that secure
the Top Case to the Bottom Case.
3.
Turn the unit right side up and gently pry the Top Case from the Bottom Case. (The
Top Case may be difficult to remove due to foam insulation. Be very careful.)
Use a 9/64” Allen wrench to
remove the socket-head
screws.
nCPAP driver with Top Case removed.