nCPAP Driver Service Manual
2008, Cardinal Health
Page 17 of 81
CAUTION
The correct AC adapter must be used with the driver. Use of an incorrect
AC adapter can damage the internal circuits of the driver.
CAUTION
Although this device meets the requirements of current EMC/RFI
legislation, this does not guarantee immunity from all sources of radiated
energy. Some mobile telephones and other products containing radio
transmitting components may cause a malfunction of the device and
should not be used in the vicinity of the device.
CAUTION
Using the driver without the proper filters can allow dirt or moisture to
enter the system. This will damage the electronic blender, leaving the unit
non-operational.
CAUTION
Ensure that the driver is unplugged from the AC Adaptor and that the
power switch is in the “Off” position before cleaning.
The driver will be damaged if any part of it is immersed in liquid or gas, or
steam sterilized.
Important: Do not spray down the unit with water or cleaning solutions.
CAUTION
This assembly contains electronic components that are ESD sensitive.
ESD control and handling measures should be followed while performing
this process.