nCPAP Driver Service Manual
2008, Cardinal Health
Page 18 of 81
CAUTION
Federal (USA) law restricts this device to sale by or on the order of a
physician or licensed practitioner.
CAUTION
Do not block or occlude flow from exiting the Oxygen Sensor Bleed Port as
this may damage the internal oxygen sensor or create internal leaks.
CAUTION
Any liquid introduced into the gas supplies will cause malfunction of this
driver and the equipment connected to it.
Note:
Always disconnect the machine from facility power, oxygen and medical air
before performing any maintenance procedures.
Note:
To enhance patient safety and comfort, use only with heated humidifiers
and heated wire circuits. The recommended setting is 37 ºC, 100% RH.
Note:
To obtain maximum battery life, it is recommended that the driver remain
connected to an external power source even when it is not in use.
Note:
The driver must have both gases properly connected and within the
specified pressure range in order to pass the initial startup sequence. This
will enable the driver to perform a valid calibration and ensure accurate
FiO
2
delivery.