70-00569-02 G
Powerheart® G5 Automated External Defibrillator
C-4
ECG Analysis Algorithm and Rescue Waveform
Waveform and energy levels for adult defibrillation pads
Table C-1: Ultra-low variable energy waveform (all values are typical)
Phase 1
Phase 2
Patient’s Impedance
(Ohms)
Voltage
(Volts)
Duration
(ms)
Voltage
(Volts)
Duration
(ms)
Energy
(Joules)
25
1,412
3.25
743
3.2
146-197
50
1,426
4.50
907
3.2
128-172
75
1,431
5.75
968
3.2
116-156
100
1,433
7.00
1,000
3.2
108-144
125
1,435
8.25
1,019
3.2
102-136
150
1,436
9.50
1,031
3.2
97-130
175
1,437
10.75
1,038
3.2
94-126
Table C-2: Low variable energy waveform (all values are typical)
Phase 1
Phase 2
Patient’s Impedance
(Ohms)
Voltage
(Volts)
Duration
(ms)
Voltage
(Volts)
Duration
(ms)
Energy
(Joules)
25
1,631
3.25
858
3.2
195-263
50
1,647
4.50
1,047
3.2
170-230
75
1,653
5.75
1,118
3.2
154-208
100
1,655
7.00
1,155
3.2
143-193
125
1,657
8.25
1,176
3.2
135-182
150
1,658
9.50
1,190
3.2
129-174
175
1,659
10.75
1,199
3.2
125-168
Содержание Powerheart G5
Страница 1: ...User Guide 70 00569 02 G Cardiac Science Powerheart G5 Automated External Defibrillator...
Страница 2: ......
Страница 3: ...USER GUIDE POWERHEART G5 AUTOMATED EXTERNAL DEFIBRILLATOR 70 00569 02 G...
Страница 8: ...Contents 70 00569 02 G vi Contents...
Страница 14: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 1 6 About the AED...
Страница 22: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 2 8 Steps of a Rescue...
Страница 36: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 4 4 AED Features...
Страница 78: ...70 00569 02 G Powerheart G5 Automated External Defibrillator C 8 ECG Analysis Algorithm and Rescue Waveform...
Страница 93: ......
Страница 94: ......
Страница 95: ......