70-00569-02 G
Powerheart® G5 Automated External Defibrillator
6-2
Product Care
Periodical maintenance
Periodically, perform the following tests.
Check the colour of the Rescue Ready® indicator.
Check that the battery has adequate charge to perform a rescue:
1.
Open the lid of the AED.
2.
If the battery indicator is red, replace the battery.
3.
Close the lid.
Check that the voice prompts work and the display is readable:
1.
Open the lid of the AED.
2.
Listen for the voice prompts.
3.
Additionally, check that the display shows text prompts that correspond to the audio.
4.
Close the lid. The voice prompts should stop.
5.
Verify that the Rescue Ready indicator returns to green.
If no prompts are heard or they continue after the lid is closed, the display is not readable,
or the Rescue Ready indicator remains red, there may be an issue with the AED. Contact
Cardiac Science Technical Support, or, outside the U.S., your local representative.
Check that the defibrillation pads are ready for use and that the service beep sounds:
1.
Open the lid of the AED.
2.
Disconnect the pad connector and remove the pad package.
3.
Close the lid.
If the colour is…
Do this…
Green
No action needed. The AED is ready for a rescue.
Red
Refer to
✓
✓
✓
✓
Содержание Powerheart G5
Страница 1: ...User Guide 70 00569 02 G Cardiac Science Powerheart G5 Automated External Defibrillator...
Страница 2: ......
Страница 3: ...USER GUIDE POWERHEART G5 AUTOMATED EXTERNAL DEFIBRILLATOR 70 00569 02 G...
Страница 8: ...Contents 70 00569 02 G vi Contents...
Страница 14: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 1 6 About the AED...
Страница 22: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 2 8 Steps of a Rescue...
Страница 36: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 4 4 AED Features...
Страница 78: ...70 00569 02 G Powerheart G5 Automated External Defibrillator C 8 ECG Analysis Algorithm and Rescue Waveform...
Страница 93: ......
Страница 94: ......
Страница 95: ......