Powerheart® G5 Automated External Defibrillator
70-00569-02 G
D-5
Guidance and manufacturer’s declaration—electromagnetic immunity
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.5 GHz
10 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the AED, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
d = 1.2
80 MHz to 800 MHz
d = 2.3
800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m)
b
.
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
c
should be less than the compliance
level in each frequency range.
d
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment—
guidance
P
P
Содержание Powerheart G5
Страница 1: ...User Guide 70 00569 02 G Cardiac Science Powerheart G5 Automated External Defibrillator...
Страница 2: ......
Страница 3: ...USER GUIDE POWERHEART G5 AUTOMATED EXTERNAL DEFIBRILLATOR 70 00569 02 G...
Страница 8: ...Contents 70 00569 02 G vi Contents...
Страница 14: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 1 6 About the AED...
Страница 22: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 2 8 Steps of a Rescue...
Страница 36: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 4 4 AED Features...
Страница 78: ...70 00569 02 G Powerheart G5 Automated External Defibrillator C 8 ECG Analysis Algorithm and Rescue Waveform...
Страница 93: ......
Страница 94: ......
Страница 95: ......