70-00969-01 F
Powerheart® AED G3 Pro 9300P
3-12
How to Perform a Rescue
6.
After pressing the shock button or 30 seconds elapses without the
shock button being pressed, the AED reverts to AED MODE
and prompt, “START CPR”. Follow the voice prompts. If
“Remain in Manual Mode” has been enabled, the device remains
in Manual Mode.
7.
To re-enter manual mode, press the
MANUAL
button ONCE.
Note:
Should the rescuer go into manual mode and decide that
AED mode is more appropriate, the AED will revert back to AED
mode 30 seconds after charging is complete. The seconds count
down on the display. If
Remain in Manual Mode
has been
enabled, the device will remain in Manual Mode.
Exiting Manual mode
By default, the device will return to AED mode after you perform any
of these actions:
◆
Press the Shock button
◆
Allow 30 seconds to elapse without pressing the shock button
◆
Close the AED lid momentarily
◆
Remove the battery momentarily
◆
Attach the optional 3-lead ECG monitoring cable
◆
Disconnect the pads from the AED
◆
Remove the pads from the patient
Exiting Manual mode when Remain in Manual Mode
is Enabled
If
Remain in Manual Mode
is enabled, the device will return to AED
mode after you perform any of these actions:
◆
Close the AED lid momentarily
◆
Remove the battery momentarily
◆
Attach the optional 3-lead ECG monitoring cable (upon
reattaching the defibrillation pads the AED will be in manual
mode)