Powerheart® AED G3 Pro 9300P
70-00969-01 F
3-3
Warnings and cautions
The following cautions must be observed to prevent problems during
the rescue.
!
DANGER! Fire and Explosion Hazard
To avoid possible fire or explosion hazard, do not operate the AED:
•
In the presence of flammable gases
•
In the presence of concentrated oxygen
•
In a hyperbaric chamber
!
WARNING! Shock Hazard and Possible Equipment Damage
Defibrillation shock current flowing through unwanted pathways
is potentially a serious electrical shock hazard and potential
damage to the equipment. To avoid this hazard during
defibrillation abide by all of the following:
•
Do not touch the patient, unless performance of CPR is
indicated
•
Do not touch metal objects in contact with the patient
•
Keep defibrillation pads clear of other pads or metal parts
in contact with patient
•
Disconnect all non-defibrillator proof equipment from
the patient before defibrillation
!
WARNING! Do not reuse pads.
Used pads may not adhere properly to the patient. Improper pad
adhesion may result in skin burns. Improper pad adhesion may result
in improper AED performance. Used pads may cause patient-to-
patient contamination.
!
Caution: Use only Cardiac Science Approved Equipment
Using batteries, pads, cables, or optional equipment other than those
approved by Cardiac Science may cause the AED to function
improperly during a rescue.
!
Caution: Possible Improper AED Performance
Using pads that are damaged or expired may result in improper AED
performance.