TM 55/65 Treadmill 4.0
70-00297-01 A
B-3
Site Requirements
Site Requirements
The treadmill requires a dedicated AC power line (refer to “Power Requirements” on
page -3). To ensure electrical safety, the treadmill is equipped with a three-wire power cord
and three-pronged plug. To maintain ground reliability, the plug must be connected to an
equivalent receptacle.
!
Caution. Operate in conditioned environment.
The treadmill is designed to operate in a typical clinical environment with adequate heat
dissipation (1850 Watts maximum).
Place the treadmill on a flat surface, free of moisture and debris. Maintain a minimum
clearance of 1.5 feet (0.5 m) on each side and 6 feet at the rear.
If the preamp for the Q-Stress system is to be attached to the handrail, maintain a
clearance of 1.5 feet (0.5 m) at the front of the treadmill.