Programming 

12 

MC18-GB – v1.04 

 

5.2 

Running a Program 

If using the 3216P5 controller Press Scroll 

3

 

to display 

PROG

 <

PROGRAM NUMBER

> select the 

required Program Number before running a Program. 
 

Table below shows the key presses to run a program 

Operation Action 

Indication 

To RUN a program  Press and quickly 

release 

S

 + 

T

 

Indicator – RUN = On 
Scrolling Display – 

Current Program state

 

To HOLD a 
program 

Press and quickly 
release 

S

 + 

T

 

Indicator – RUN = Flashing 
Scrolling Display – Program Hold 

To RESET a 
program 

Press and Hold  

S

 + 

T

 

for more than 

1 second 

Indicator – RUN = Off 
Scrolling Display - 

None

 

 

Program Ended 

Indicator – RUN = Off 
Scrolling Text – Program End 

To RESET a 
program after it 
has completed 

Press and Hold 

S

 + 

T

 

for more than 

1 second 
Or 
Press and quickly 
release 
Ack  

3

Indicator – RUN = Off 
Scrolling Display - 

None

 

 

5.3 Program 

Status 

 
While the program is running in level 1 or level 2, the Home Display shows two values at 
any one time: 
 
Process Value 
The upper display shows the measured temperature of the furnace (PV). 
 
PSP, Segment Type and Number 
The lower display continually alternates between the programs current set value (Program 
SP = PSP) and scrolling text, indicating the current status of the program whether RAMP or 
DWELLING followed by the segment number. 
 
Additional information can be obtained using the scroll 

3

 key while the program is running. 

 
Working Output Power 
From the home display, press Scroll 

3

 

until the display shows 

WRK.OP

 

<

WORKING OUTPUT 

POWER

>

 This shows the power being used as a percentage.  

Содержание 3216 P1

Страница 1: ...MC18 GB v1 04 Operating Instructions Temperature Controller Type 3216 P1 P5 English...

Страница 2: ...2 MC18 GB v1 04...

Страница 3: ...LER 8 4 1 MAXIMUM OUTPUT POWER SETTING 8 4 2 CUSTOMER ID 8 4 3 UNITS 8 4 4 LANGUAGE 8 4 5 SCROLLING TEXT 8 4 6 CUSTOMER CALIBRATION 9 4 7 HOLDBACK 10 5 PROGRAMMING 10 5 1 CREATING A PROGRAM 10 5 2 RUN...

Страница 4: ...controller which uses PID algorithms to give excellent temperature control This controller can store and run a single program of 8 ramp dwell segments The 3216P1 can also be used as a simple temperatu...

Страница 5: ...n alarm if activated Reset a program after the program has ended Arrow Keys S T The arrow keys are used individually to adjust the selected parameters and in combination to run a program Note If a par...

Страница 6: ...The scrolling text can be interrupted at any time by a single press of any of the buttons but will not scroll again until the parameter is returned to In this manual the mnemonic will be shown first f...

Страница 7: ...cted the display reverts to the home display See Section 2 1 Table showing parameters accessible in level 1 and Level 2 Operator LEVEL 1 Supervisor LEVEL 2 Home display Home Display Programming Progra...

Страница 8: ...wer setting depends on the supply voltage refer to the product manual for details 4 2 Customer ID A furnace or oven identification number can be entered if required This maybe used to identify one of...

Страница 9: ...he wrong pass code is entered the display will revert to zero pass code until the correct pass code is entered When the correct pass code is entered and PNT LO Adjust Low Point is displayed Use Up S D...

Страница 10: ...then continue to control again The hold back will only apply once per Segment therefore when control has been re established the hold back will not apply again to that segment even if the furnace or o...

Страница 11: ...play shows T SP 1 Target SP 1 Enter the temperature that you want the unit to ramp up to using Ramp rate 1 If Ramp Rate 1 has been set to OFF The unit will Heat Cool directly to the Target SP Dwell Ti...

Страница 12: ...a program after it has completed Press and Hold S T for more than 1 second Or Press and quickly release Ack 3 Indicator RUN Off Scrolling Display None 5 3 Program Status While the program is running i...

Страница 13: ...a power failure while the program is running and the power is subsequently restored the scrolling text will read POWER FAIL PROGRAM RESET To acknowledge this message and to reset the program Press ACK...

Страница 14: ...ay shows T SP3 TARGET SP 3 Select 600 11 Press Scroll until the display shows DWEL 3 DWELL TIME 3 Select 60 12 Press Scroll until the display shows RMP 4 RAMP RATE 4 Select 3 13 Press Scroll until the...

Страница 15: ...ed but may not be necessary The cable is usually 9 pin at the furnace end and 9 pin at the computer but other alternatives are shown in parentheses Furnace end of cable RS232 Cable furnace to PC Compu...

Страница 16: ...The purpose of the 2 amp fuse is to guard against the connection of a mains supply to the panel plug The instrument configuration and parameters available to the operator depend on the customer requir...

Страница 17: ...t sp1 A4 dwel 1 op hi rmp 8 LD AMP t sp8 LK AMP dwel 8 LD ALARM LD AMP If Configured LK ALARM ID HC ALARM ADDR ID UNITS CAL P PNT LO OFS LO PNT HI OFS HI Through to If Configured Through to Multi prog...

Страница 18: ...Thermocouple The temperature measuring device used in the furnace or oven PID Proportional Integral Derivative the control system used by the controller Overtemperature O T The condition which a furn...

Страница 19: ...Notes MC18 GB v1 04...

Страница 20: ...hermal Engineering Services Telephone UK 0845 3308035 Int 44 1433 623335 Fax UK 0845 3308036 Int 44 1433 623336 Email service thermalserve com Carbolite Parsons Lane Hope Hope Valley S33 6RB England T...

Отзывы: