Programming
11
MC18-GB – v1.04
Ramp Units
Press scroll
3
until the display shows
RAMP.U
<
Ramp Units
>
: Use the Up
S
Down
T
to
select the Ramp Units of
Hour, Min or Seconds
.
Dwell Units
Press Scroll
3
until the display shows
DWEL.U
<
Dwell Units
>
: Use the Up
S
Down
T
to
select the Dwell Units of
Hour or Min
.
Holdback (See section 4.7)
Press Scroll
3
until the display shows
H.BACK
<
Program Hold Back
>
: If a Holdback value is
required, use the Up
S
Down
T
to enter the required value or select “
OFF
” if no holdback is
required.
Ramp Rate
Press Scroll
3
until the display shows
RMP.1
<
Ramp Rate 1
>
: Using the Up
S
Down
T
enter the
value for the first ramp rate depending on the Ramp Units selected. If the Ramp
segment is not required select “
OFF
”.
Target Setpoint
Press Scroll
3
until the display shows
T.SP 1
<
Target SP 1
>
: Enter the temperature that
you want the unit to ramp up to using “Ramp rate 1”.
If “
Ramp Rate 1
” has been set to “OFF”. The unit will Heat/Cool directly to the Target SP.
Dwell Time
Press Scroll
3
until the display shows
DWELl. 1
<
DWELL TIME 1
>
: Enter the time to dwell
at “Target SP 1”. If the dwell segment is not required, use the Down
T
to select
OFF
, which
is below the Zero value.
This Process is repeated for each of the 8 segments of the program.
If not all the segments are used for a program, the Ramp & Dwell of each of the subsequent
Segments should be set to
OFF
.
A program will end in one of two ways, either revert to the control Set Point or dwell at the
temperature set in the last segment used. When a program finishes on a dwell and the dwell
time expires the temperature will revert to the control setpoint.
Note
: Before running a program ensure that the control set point is set to Zero to avoid
unexpected heating at the end of the program.