Programming 

10 

MC18-GB – v1.04 

 

Single Offset 

If a constant offset is required across the temperature range, set the required “High Point” 
(

PNT.HI

) and “Low Point” (

PNT.LO

) to the required values (not the same), then set the “low 

Offset” (

OFS.LO

) and “High Offset” (

OFS.HI

) to the same value. 

 

Caution! - Do not make 

PNT.LO

 and 

PNT.HI

 the same value as the controller will 

not work correctly and could cause your furnace or oven to overheat. 

 

4.7 Holdback 

 
If the temperature ramp rate of the program is quicker than the furnace or oven can achieve, 
the program will wait until the temperature of the furnace or oven catches up. 
e.g. If a Holdback value of 10 is set and the program is set to ramp to a setpoint of 600°C, 
the program will reach 600°C, then go into an hold state; the hold indicator will light until the 
furnace or oven temperature reaches 590°C, the program will then continue to control 
again. 
The hold back will only apply once per Segment, therefore when control has been re-
established, the hold back will not apply again to that segment, even if the furnace or oven 
temperature go outside the hold back band.  
 
Holdback can only be accessed in supervisor level (level 2) by scrolling with the Scroll 

3

 

until the display shows

 

H.BACK

 <

PROGRAM HOLD BACK

>: Use the Up 

S

 Down 

T

 to select the 

required Hold Back value. 
If a multi programmer is used, each program can have its own Holdback value assigned to 
it. 
 

Note:

 

When a Holdback is set, each segment used must have a Ramp Rate assigned 

to it, in order for it to be recognised by the program.

 

 

5 Programming 

 

5.1 

Creating a Program 

Programs can be created in level 1 or level 2 of the 3216 P1 and 3216P5.  Each program 
contains 8 Ramp/Dwell pairs. 
 

Note: 

A currently active program cannot be altered. Go into reset mode before starting to create or 
modify a program 
 
 
Program Number (3216 P5 Only)

 

Press scroll 

3

 

until the display shows 

PROG

 

<

PROGRAM NUMBER

.

>

 Select the program 

number.  
 

Содержание 3216 P1

Страница 1: ...MC18 GB v1 04 Operating Instructions Temperature Controller Type 3216 P1 P5 English...

Страница 2: ...2 MC18 GB v1 04...

Страница 3: ...LER 8 4 1 MAXIMUM OUTPUT POWER SETTING 8 4 2 CUSTOMER ID 8 4 3 UNITS 8 4 4 LANGUAGE 8 4 5 SCROLLING TEXT 8 4 6 CUSTOMER CALIBRATION 9 4 7 HOLDBACK 10 5 PROGRAMMING 10 5 1 CREATING A PROGRAM 10 5 2 RUN...

Страница 4: ...controller which uses PID algorithms to give excellent temperature control This controller can store and run a single program of 8 ramp dwell segments The 3216P1 can also be used as a simple temperatu...

Страница 5: ...n alarm if activated Reset a program after the program has ended Arrow Keys S T The arrow keys are used individually to adjust the selected parameters and in combination to run a program Note If a par...

Страница 6: ...The scrolling text can be interrupted at any time by a single press of any of the buttons but will not scroll again until the parameter is returned to In this manual the mnemonic will be shown first f...

Страница 7: ...cted the display reverts to the home display See Section 2 1 Table showing parameters accessible in level 1 and Level 2 Operator LEVEL 1 Supervisor LEVEL 2 Home display Home Display Programming Progra...

Страница 8: ...wer setting depends on the supply voltage refer to the product manual for details 4 2 Customer ID A furnace or oven identification number can be entered if required This maybe used to identify one of...

Страница 9: ...he wrong pass code is entered the display will revert to zero pass code until the correct pass code is entered When the correct pass code is entered and PNT LO Adjust Low Point is displayed Use Up S D...

Страница 10: ...then continue to control again The hold back will only apply once per Segment therefore when control has been re established the hold back will not apply again to that segment even if the furnace or o...

Страница 11: ...play shows T SP 1 Target SP 1 Enter the temperature that you want the unit to ramp up to using Ramp rate 1 If Ramp Rate 1 has been set to OFF The unit will Heat Cool directly to the Target SP Dwell Ti...

Страница 12: ...a program after it has completed Press and Hold S T for more than 1 second Or Press and quickly release Ack 3 Indicator RUN Off Scrolling Display None 5 3 Program Status While the program is running i...

Страница 13: ...a power failure while the program is running and the power is subsequently restored the scrolling text will read POWER FAIL PROGRAM RESET To acknowledge this message and to reset the program Press ACK...

Страница 14: ...ay shows T SP3 TARGET SP 3 Select 600 11 Press Scroll until the display shows DWEL 3 DWELL TIME 3 Select 60 12 Press Scroll until the display shows RMP 4 RAMP RATE 4 Select 3 13 Press Scroll until the...

Страница 15: ...ed but may not be necessary The cable is usually 9 pin at the furnace end and 9 pin at the computer but other alternatives are shown in parentheses Furnace end of cable RS232 Cable furnace to PC Compu...

Страница 16: ...The purpose of the 2 amp fuse is to guard against the connection of a mains supply to the panel plug The instrument configuration and parameters available to the operator depend on the customer requir...

Страница 17: ...t sp1 A4 dwel 1 op hi rmp 8 LD AMP t sp8 LK AMP dwel 8 LD ALARM LD AMP If Configured LK ALARM ID HC ALARM ADDR ID UNITS CAL P PNT LO OFS LO PNT HI OFS HI Through to If Configured Through to Multi prog...

Страница 18: ...Thermocouple The temperature measuring device used in the furnace or oven PID Proportional Integral Derivative the control system used by the controller Overtemperature O T The condition which a furn...

Страница 19: ...Notes MC18 GB v1 04...

Страница 20: ...hermal Engineering Services Telephone UK 0845 3308035 Int 44 1433 623335 Fax UK 0845 3308036 Int 44 1433 623336 Email service thermalserve com Carbolite Parsons Lane Hope Hope Valley S33 6RB England T...

Отзывы: