-29-
Instruction Manual
22. Press the "
ON"
switch, select program, "(00)," and press the
"START SANITIZE"
switch. Allow the machine
to operate to completion of PROGRAM STEP
75
and press
"OFF"
switch (No problems will occur if machine
operates past step
75
before turning off). This action clears the water from the uptake hose and replaces it with
RENALIN
®
100 Cold Sterilant.
23. Press the "
ON"
switch, select program, "(00)," and press the
"START SANITIZE"
switch. Allow the machine
to operate to PROGRAM STEP
83
. When the machine enters step
83
, press the
"START PROCESS"
switch.
The machine will then enter step
84
and after a ten minute hold time will proceed to PROGRAM STEP
97
.
The
"SANITIZE COMPLETE"
message will reappear.
24. Your RENATRON
®
II 100 Series station is now ready for normal use.
Note – Formula 409
®
is a registered trademark of the Clorox Co., Oakland, CA. To order use (PN 93250-206).
MEDIVATORS International Distributors should contact MEDIVATORS to obtain this product if not available
locally.
Formula 409 Service Cleaning Procedure for RENATRON Stations - RS-8330
1. Remove the uptake tube from the RENALIN 100 container and make sure excess
Renalin
100 drains back into
the container. Cover the open RENALIN 100 container.
2.
Place the uptake tube into a container of purified water (one gallon or more).
3. Press the
"ON"
switch, and select program "00." Connect external hoses in sanitize configuration.
4. Press the
"START SANITIZE"
switch. Allow the machine to operate to completion of PROGRAM STEP
75
and press the
"OFF
" switch. (No problems will occur if machine operates past step 75 before turning off).
5. Press the
"ON"
switch and repeat procedure steps 3 and 4 above. This action clears RENALIN 100 from the
uptake hose and replaces it with purified water.
6.
Remove the uptake tube from the purified water and allow excess water to drain out of the uptake tube.
7. Place the uptake tube into container of Formula 409. Do not use a substitute chemical as it may damage the
materials in the machine.
8. Press the
"ON"
switch and repeat procedure steps 3 and 4 above. This action clears the water from the uptake
hose and replaces it with Formula 409.
9. Press
"ON,"
select program, "(00)," press the
"START SANITIZE"
switch. Allow the machine to continue
operating to the
"SANITIZE COMPLETE"
message in PROGRAM STEP
83
.
10. Press the "OFF" switch and turn the water supply to the machine off. Leave the machine overnight. Wait at
least 8 hours.
11. Remove the uptake tube from the container of Formula 409 and make sure any excess Formula 409 drains from
tube back into container. Place the uptake tube in a container of purified water (one gallon or more). Turn the
water supply on.
12. Press
"ON,"
select program, "(00)," press the
"START SANITIZE"
switch. Allow the machine to operate to
completion of PROGRAM STEP
75
and press the
"OFF"
switch. (No problems will occur if machine operates
past step 75 before turning off.) This action clears the Formula 409 from the uptake hose and replaces it with water.
13. Press
"ON,"
select program, "(00)," press the
"START SANITIZE"
switch. Allow the machine to operate
to PROGRAM STEP
83
, and when machine completes step 83, press the
"START PROCESS"
switch. The
machine will then enter step 84 and, after a ten minute hold period in step 84, will proceed with flushing and
draining. The flushing and draining will continue until machine enters PROGRAM STEP
97
and the
"SANITIZE
COMPLETE"
message reappears.
14. Press the
"OFF"
switch when th
e "SANITIZE COMPLETE"
message appears in PROGRAM STEP
97
.
15.
Remove the uptake tube from the container of purified water and allow excess water to drain from uptake tube.
16. Place the uptake tube in a container of unexpired RENALIN 100 solution.
17. Press the
"ON
" switch, select program, "(00)," and press the
"START SANITIZE"
switch. Allow the machine
to operate to completion of PROGRAM STEP
75
and press
"OFF"
switch (No problems will occur if machine
operates past step 75 before turning off). This action clears the water from the uptake hose and replaces it with
RENALIN 100.
18. Press the
"ON"
switch, select program, "(00)," and press the
"START SANITIZE"
switch. Allow the machine
to operate to PROGRAM STEP
83
. When the machine enters step 83, press the "
START PROCESS"
switch.
The machine will then enter step 84 and after a ten minute hold time will proceed to PROGRAM STEP
97
. The
"SANITIZE COMPLETE"
message will reappear.
19. Your RENATRON II 100 Series station is now ready for normal use.