-26-
Instruction Manual
NOTE – The new calibration values are not saved until you complete step 10 and exit the calibration mode.
11. Press the
"MUTE"
and
"RESET"
switches together until program "(00)" is displayed.
12. Press the
"HOLD TO SET"
switch while adjusting the
"SET"
knob to display a volume greater than 200 in
the
"VOLUME"
window.
13 Press the
"START PROCESS"
switch. A light will appear in the corner of the
"START PROCESS"
switch.
14. Allow the RENATRON
®
II 100 Series station to continue operating until it enters
"PROGRAM STEP 8"
.
15. Press the
"OFF"
switch.
16. Verify that the liquid level in the tank is at the zero reference line (lower tank scribe line).
17. If the liquid level is above or below the zero reference line repeat steps 3-16.
18. Press the
"ON"
switch.
19. Press the
"MUTE"
AND
"RESET"
switches together until program "(00)" is displayed.
20. Press the
"HOLD TO SET"
switch while adjusting the
"SET"
knob to display a volume greater than 200 in
the
"VOLUME"
window.
21. Press the
"START PROCESS"
switch. A light will appear in the corner of the
"START PROCESS"
switch.
22. Allow the RENATRON
II 100 Series station to continue operating until it enters
"PROGRAM STEP 14"
.
23. Press the
"OFF"
switch.
24. Verify that the liquid level in the tank is at the 507 reference line (upper tank scribe line).
25. If the liquid level is above or below the 507 reference line repeat steps 3-24.
26. Press the
"ON"
switch.
27. Press the
"MUTE"
AND
"RESET"
switches together until program "(00)" is displayed.
28. Press the
"HOLD TO SET"
switch while adjusting the
"SET"
knob to display "a volume greater than 200"
in the
"VOLUME"
window.
29. Press the
"START PROCESS"
switch. A light will appear in the corner of the
"START PROCESS"
switch.
30. Allow the RENATRON II 100 Series station to continue operating until the
"Volume Fail"
alarm has occurred
in "
PROGRAM STEP 35"
. Push the
"MUTE"
switch to silence the audio alarm.
DO NOT PRESS THE
"RESET" SWITCH.
31. The displayed RENATRON
II 100 station volume value must be within ±3 ml of the Calibrated Volume Cell.
If the volume is not within ±3 ml, press the
"RESET"
switch and repeat steps 4 - 30.
32. Replace the RENATRON II 100 Series station cover, securing it with the two screws on each side.
CAUTION
– THE COVER SHOULD ALWAYS BE ON THE RENATRON II 100 SERIES STATION FOR
NORMAL OPERATION.
Note – If the RENATRON
II 100 Series station will not calibrate, please call MEDIVATORS Renal Technical
Support at 1-800-328-3324 for assistance.