-20-

Instruction Manual

CAUTION

 – IT IS THE SOLE RESPONSIBILITY OF THE PRESCRIBING PHYSICIAN TO DETERMINE 

THE ACCEPTABLE BLOOD COMPARTMENT REFERENCE VOLUME AND PROGRAM MODE FOR 

A PARTICULAR TYPE AND SIZE HOLLOW FIBER DIALYZER. THIS REFERENCE VOLUME AND 

PROGRAM MODE MUST BE SET FROM THE FRONT PANEL MANUALLY OR INITIALLY ENTERED 

INTO THE DATABASE TO ALLOW THE COMPUTER TO AUTOMATICALLY SET THESE VALUES BEFORE 

REPROCESSING A DIALYZER. IT IS STRONGLY RECOMMENDED THAT THE MINIMUM ACCEPTABLE 

REFERENCE VOLUME BE AT LEAST 80% OF THE PRE-PROCESS VOLUME.

CAUTION

 – IT IS IMPORTANT FOR THE OPERATOR TO VERIFY THAT THE BLOOD COMPARTMENT 

REFERENCE VOLUME SET ON THE MACHINE MANUALLY OR AUTOMATICALLY IS CORRECT FOR 

EACH DIALYZER TO BE REPROCESSED.

E. 

Reprocessing the Dialyzer on the RENATRON

®

 

  

II 100 Series station in manual operation

1. 

Push the switch labeled "ON.

" "SEL" will be displayed in the 

"VOLUME"

 display window and "

– –

will appear in the "PROGRAM STEP" display window.

2. 

Push the "RESET" switch and the "MUTE" switch at same time

 to display the proper program mode in 

the "PROGRAM STEP" display window for the dialyzer to be reprocessed, i.e. "00," "CH," "HF." 

Note  – Pre-clean (PC) mode is disabled when using the RENATRON II 100 Series station in manual operation.

3. 

Touch and hold the "HOLD TO SET" switch

.

4. 

Adjust  the  "SET"  knob  until  the  "VOLUME"  display  indicates  the  physician’s  prescribed  blood 

compartment reference volume

 for the dialyzer to be reprocessed.

5. 

Touch the "START PROCESS" switch

. The RENATRON II 100 Series station will now reprocess the 

dialyzer in the manner described in the section General Systems Operation, of this manual.

6.  When the 

"PROCESS COMPLETE"

 message appears the reprocessed dialyzer is ready to be removed from 

the RENATRON II 100 Series station.

7.   If a 

"VOLUME FAIL"

 or 

"PRESSURE FAIL"

 message appears 

press the "START PROCESS" switch 

to repeat the test which was failed

. The dialyzer may pass the test when retested by the RENATRON II 100 

Series station.

F.  

Reprocessing the dialyzer on the RENATRON

 

II 100 Series station in automatic operation

1. 

Push switch labeled "ON.

" "SEL" will be displayed in the

 "VOLUME"

 display window and "--" will appear 

in the "PROGRAM STEP" display window.

2.  In the automatic reprocessing screen of the RENALOG

®

  RM Program, when the green light is illuminated 

on interface module, 

scan the bar code label on the dialyzer to be reprocessed, scan the bar code label 

on the RENATRON II 100 Series station which corresponds to the cycle that is to be run and scan your 

authentication label

. The order of the labels being scanned does not make a difference.

Note – The two cycles available are the complete reprocessing cycle and  the pre-clean cycle. To select a complete 

reprocessing cycle, scan the barcode label on the RENATRON II 100 Series station, the dialyzer's bar coded label. 

After the third label, your authentication label is scanned, the computer will automatically set the prescribed 

blood compartment reference volume and correct program mode for the dialyzer to be reprocessed. To select the 

pre-clean option, scan the bar code labeled pre-clean, the dialyzer's bar code label and your authentication label. 

Note – Maximum effectiveness of the pre-clean mode is attained in two hours. During this period, if the dialyzer is 

not removed from the RENATRON II 100 Series, the machine will vent the dialyzer every 10 seconds to prevent 

pressure build up. One label will be printed at the end of the pre-clean cycle warning you that the dialyzer has 

been pre-cleaned only and must go through the complete reprocessing cycle before being placed in storage.

CAUTION

 – THE PRE-CLEAN MODE IS INTENDED TO AUTOMATICALLY STOP THE RENATRON II 100 

SERIES STATION AFTER THE RENALIN 100 CLEANING PHASE. THIS ACTION ALLOWS THE OPERATOR 

TO REMOVE THE DIALYZER AND SET IT ASIDE FOR AN ADDITIONAL SOAK/DWELL CLEANING PERIOD. 

THIS PROCESS HELPS CLEAN DIFFICULT DIALYZERS. PRECLEANED DIALYZERS MUST UNDERGO A 

FULL REPROCESSING CYCLE BEFORE USE.

3. 

Touch the "START PROCESS" switch

. The RENATRON II 100 Series station will now reprocess the 

dialyzer in the manner described in the section General Systems Operation, of this manual.

4. 

Push the "RESET" switch on the RENATRON II 100 Series station

 after a 

"PROCESS COMPLETE"

"VOLUME FAIL"

, or

 "PRESSURE FAIL"

 message has appeared. The 

"PROGRAM STEP"

 display will 

indicate step 

98

. Within a few seconds, the computer will read and reset the RENATRON II 100 Series station 

and the last program mode used will be displayed in the 

"PROGRAM STEP"

 display. After the computer has 

read the data, labels are printed out indicating the dialyzer's status. Place one label on the dialyzer. Additional 

labels may be printed if needed.

Содержание MEDIVATORS renatron II 100 Series

Страница 1: ...100 Series Dialyzer Reprocessing System INSTRUCTION MANUAL MODEL RS 8330 RS 8335 100 Series...

Страница 2: ...IN PERASSAY ACTRIL and ON DEMAND are registered trademarks of Medivators Inc Formula 409 is a registered trademark of the Clorox Company 50096 725 Rev D 2015 Medivators Inc Printed and Assembled in th...

Страница 3: ...tions 12 8 Computer Installation Instructions 16 9 Daily Machine Sanitize Procedure 17 10 Operating Procedures 17 11 Bar Code Scanner Procedures 23 12 Storage and Handling for Reprocessed Dialyzers 24...

Страница 4: ...and delivery of proportioned RENALIN 100 Cold Sterilant solution to hollow fiber dialyzers It is also indicated for record keeping of these functions The RENATRON II 100 Series system is indicated fo...

Страница 5: ...TION THE OPERATOR SHOULD BE FAMILIAR WITH THE SAFE HANDLING TECHNIQUES FOR THE MACHINE AND THE RENALIN 100 COLD STERILANT AND WITH THE RECOMMENDED FIRSTAID PROCEDURES IN CASE OF CONTACT WITH OR EXPOSU...

Страница 6: ...tem for rinsing cleaning and testing of hollow fiber dialyzers and for delivery of Proportioned RENALIN 100 Cold Sterilant solution to the dialyzers The RENATRON II 100 Series System can also perform...

Страница 7: ...dialyzer measured priming volume may differ from advertised volume by as much as 20 It is recommended that the operator measure new dialyzer volume with a known calibrated RENATRONII 100 Series to det...

Страница 8: ...nstruction Acrylic PVC ABS Polypropylene Delrin Ethylene Propylene Viton Rubber Stainless Steel B Electrical Leakage Current risk current Each RENATRON II 100 Series is allowed a leakage current less...

Страница 9: ...3 cm Weight 3 2 lbs 1 5 kg 4 Report Printer approximately Width 17 7 inches 50 cm Depth 14 6 inches 37 cm Height 5 7 inches 15 cm Weight 11 7 lbs 5 3 kg 5 Monitor LCD approximately Width 14 8 inches 3...

Страница 10: ...e RENATRON station will perform a header leak test between program Step 2 and 3 If a Pressure Fail message appears in Step H 2 verify that the blood port connections are secure and that the dialyzer h...

Страница 11: ...display to SEL Reset must be used at the end of a cycle or in an alarm state locked out or disabled at all other times to guarantee automatic transfer of reprocessing data to the computer 7 DISPLAY sw...

Страница 12: ...er to the respective manuals for detailed information about the computer monitor and printer controls and indicators IMPORTANT TO ISOLATE THE STATION FROM THE ELECTRICAL SUPPLY YOU MUST DISCONNECT THE...

Страница 13: ...13 Instruction Manual Figure 1 RS 8335...

Страница 14: ...14 Instruction Manual Figure 2 RS 8330...

Страница 15: ...he water pressure gauge block to the back of the drip tray using the screws provided Be absolutely certain that no water leaks exist and all connections are secure G Connect the dialysate line quick d...

Страница 16: ...the computer manufacturer s manual for instructions for connecting the monitor keyboard and mouse G Power on sequence of computer system 1 Printer s 2 Monitor 3 Computer 4 Interface module 5 RENATRON...

Страница 17: ...UP THE RENALOG RM PROGRAM BEFORE NORMAL ONLINE USE OF THE SYSTEM IS ATTEMPTED THERE IS A SEPARATE RENALOG RM INSTRUCTION MANUAL P N 50090 859 A General Considerations and Cautions CAUTION MEDIVATORS S...

Страница 18: ...100 SERIES COMM PORT 2 RENATRON II 100 SERIES COMM PORT 5 RENATRON II 100 SERIES COMM PORT 3 RENATRON II 100 SERIES COMM PORT 6 RENATRON II 100 SERIES COMM PORT Port 6 Port 5 Port 4 SCANNER Port 3 Por...

Страница 19: ...rocessing Management System follow the calibration procedure outlined in the RENALOG RM User Guide PN 50090 859 C Remove the Dialyzer from the Extracorporeal Blood Circuit using standard facility prot...

Страница 20: ...screen of the RENALOG RM Program when the green light is illuminated on interface module scan the bar code label on the dialyzer to be reprocessed scan the bar code label on the RENATRON II 100 Serie...

Страница 21: ...e a supply of disinfected port caps Disinfection of the dialysate port caps blood port caps and the reprocessing connectors from the dialyzer can be achieved by one of the following ways a Place the c...

Страница 22: ...ble waterproof or permanent ink or other indelible marking systems to label the dialyzer is recommended when the RENATRON II 100 Series station is operated without RENALOG RM In the automaticmode thei...

Страница 23: ...bar code and the gun should be held 6 to 9 inches from the bar code 2 With a straight vertical motion move the scanner down over the label so that the red beam passes over the bar code The illustratio...

Страница 24: ...TION EXPOSING DIALYZERS TO CYCLIC TEMPERATURE DIFFERENCES DURING ANY PART OF THE REUSE PROCEDURE MAY RESULT IN INCREASING OR INITIATING PRESSURE BUILDUP IN THE DIALYZER DURING STORAGE IN THESE CONDITI...

Страница 25: ...ss ON switch SEL and will be displayed 4 Press and hold the SETUP switch S3 on the microprocessor board for 4 seconds until you hear the second chirp The PROGRAM STEP will display c1 and CAL will be d...

Страница 26: ...inue operating until it enters PROGRAM STEP 14 23 Press the OFF switch 24 Verify that the liquid level in the tank is at the 507 reference line upper tank scribe line 25 If the liquid level is above o...

Страница 27: ...TS DO NOT USE BLEACH C Formula 409 Service Cleaning Procedure The internal fluid pathways must be cleaned periodically typically once every two weeks or as required as outlined below RENATRONII100Seri...

Страница 28: ...ess the OFF switch and turn the water supply to the machine off Leave the machine overnight Wait at least 8 hours 14 Remove the uptake tube from the container of Formula 409 and make sure any excess F...

Страница 29: ...e hose and replaces it with Formula 409 9 Press ON select program 00 press the START SANITIZE switch Allow the machine to continue operating to the SANITIZE COMPLETE message in PROGRAM STEP 83 10 Pres...

Страница 30: ...t a sample from the calibration cell for a culture of water from the RENATRON station b After the sample has been collected press the RESET switch The RENATRON II 100 Series Station is now ready for c...

Страница 31: ...ATORS Renal Technical Support for the appropriate packing and shipping instructions Do not ship any part of the RENATRON II 100 Series system to MEDIVATORS without first notifying the Renal Technical...

Страница 32: ...Inspect Reprocessing Connections X X X X Inspect RENALIN 100 Uptake X X Tubing Assembly Inspect valve Assembly Large Small X V15 Inspect Hydraulic Compartment for X any signs of fluid leakage Test Jet...

Страница 33: ...ive Products or any parts it deems defective or ii a refund of the purchase price This will be customer s exclusive remedy for a covered defect In order to recover under the warranty the customer must...

Страница 34: ...Fuse Location Note Parts service and a service manual with drawings parts lists circuit diagrams are available on request from MEDIVATORS Renal Technical Support Explanation for Symbols Technical Supp...

Страница 35: ...ing Management Software 78398 192 RENALIN Residual Test Strips 78198 000 PERASSAY 500 Peracetic Acid Test Strips 78378 000 Dialyzer Dialysate Port Caps Red 78196 010 100 Ventable Dialyzer Dialysate Po...

Страница 36: ...729 Medivators BV Sourethweg 11 6422PC Heerlen The Netherlands Tel 31 45 5 471 471 Cantel Medical Asia Pacific Pte Ltd 1A International Business Park 05 01 Singapore 609933 Tel 65 6227 9698 Cantel Med...

Отзывы: