19
The display shows this message because of many reasons. Please, take a look at the above list with the main
possible problems:
Reasons, the display shows the message “Lo”
Advise
Temperature measurement is difficult from hair or sweating.
Ensure that there are no obstacles or moisture before
temperature measurement.
Temperature measurement is difficult from air flows or sharp
changes in environment temperature.
Ensure that there are no air flows around the area, in which
the temperature is measured, this can affect the infrared
reading.
Successive measurements of the temperature are too close
one after another and the thermometer do not have time to
restart.
Leave interval 3-5 secs. Between measurements;
recommended is pause for 15 sec.
Measurement area is away from the thermometer.
Measure the temperature on recommended distance
(around 3-5 cm, 1.2 ~2.0 inches).
SYMBOL EXPLANATION
Symbol
Explanation
Trade mark.
IEC 60417-5333, type attached part BF
IEC 60417-5031, Direct current
IP22
Protected access with fingers to dangerous parts and against vertical falling of water drops, when it is tilted up to
15 °C
Directing to the instruction/Manual.
DISPOSING: Do not throw this product with household waste. Collect separately, because special disposal of
those kind of waste is required.
SN
Specific serial number.
DECLARATION OF CONFORMITY
Manufacturer guide and declaration – electronic immunity
Easy Check is intended for using in electromagnetic environment, specified below. The client or user of JXB-183 must be
sure that the product is used in such environment.
Immunity test
IEC 60601 level test
Level of compliance
Electromagnetic environment -
guide
Electromagnetic discharge
(ESD) IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors must be wooden,
concrete or ceramic tiles. If the
floors are covered with synthetic
materials, relative humidity must
be at least 30 %.
Electrical fast transient,
explosive IEC 61000-4-4
± 2 kV за power lines
± 1 kV for input-output lines
NOT APPLICABLE
The quality of the basis power
must be this of typical trade or
hospital environment.