20
Manufacturer guide and declaration – electronic immunity
Easy Check is intended for using in electromagnetic environment, specified below. The client or user of JXB-183 must be
sure that the product is used in such environment.
Immunity test
IEC 60601 level test
Level of compliance
Electromagnetic
environment - guide
Waves
IEC 61000-4-5
± 1 kV line to line
± 2 kV line to ground
NOT APPLICABLE
Quantity of the basis power
must be this on typical trade
and hospital environment.
Voltage dips, short
interruptions and voltage
variation of the input power
lines IEC 61000-4-11
< 5% UT
(>95% dip in UT) for 0,5
cycle
40% UT
(60% dip in UT) for 5 cycle
70% UT
(30 % dip in UT) for 25
cycle
<5% UT
(>95% dip in UT) for 5 secs
NOT APPLICABLE
Mains power quality should
be that of a typical
commercial
or hospital environment. If
the
user of the “JXB-183”
requires
continued operation during
power mains interruptions, it
is recommended that the
“JXB-183”
be powered from an
uninterruptible power supply
or a battery.
Power frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Frequency magnetic fields
must be on the levels of the
typical location of the
typical trade or hospital
environment.
NOTE: UT is pressure of the alternating voltage before applying the testing level.
Guides and declarations from the manufacturer – electric immunity
Easy Check is intended for using in electromagnetic environment, аspecified below. The client or user of JXB-183 must
ensure, that it is used in such environment.
Immunity test
IEC 60601 level test
Compliance level
Electromagnetic environment- guidelines
Portable and mobile radiofrequency equipment
must not be used close to none of the parts of
“Easy Check” , including cable, except the
recommended distance, calculated from the
equation, applicable for transmitter’s frequency.
Conducted
radiofrequency
IEC 61000-4-6
3 Vrms
150 kHz for80 MHz
NOT APPLICABLE
80MHz for 800 MHz
800MHz for 2.5 GHz
Immunity test
IEC 60601 level test
Compliance level
Electromagnetic environment- guidelines
Broadcasted
radio frequency
IEC 61000-4-3
3 V/m
80 MHz to 2,5 GHz
3 V/m
Where Р is the maximum nominal power in
vats (W), according to the manufacturer’s
transmitter and “d” is the recommended
distance in meters (m).
Power of the filed from fixed RF transmitters,
as described from survey of electromagnetic
objects, “а” must be less than level of
compliance in every frequency range “b”.