![Campbell WXT520 Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/campbell/wxt520/wxt520_instruction-manual_3806653013.webp)
WXT520 Weather Transmitter
5
FIGURE 6-2. Optional bird spike device
6.4 Wiring Table
TABLE 6-1. Connections to Campbell Scientific Dataloggers
Wire Color
Wire Function
Datalogger Connection
Terminal
Brown
Power
12V
or
SW_Battery
Clear (silver)
or Red
Power ground
G
Blue
SDI-12 signal
C, U
configured for SDI-12
1
,
C1/SDI-12
, or
SDI-12
White
SDI-12 signal
C, U
configured for SDI-12
1
,
C1/SDI-12
, or
SDI-12
Green
Data ground
G
Green
Optional heater power
(see note)
12V
or
SW_Battery
Pink
Optional heater ground
(see note)
G
Grey
Not used
1
U
channels are automatically configured by the measurement instruction.
Unless special ordered, the heater will not be operational for
WXT520s purchased from Campbell Scientific. Although the
heater is not operational, the WXT520 will have a pink and yellow
wire. Do not connect the pink and yellow wire unless the heater
is operational.
NOTE