Table 8-1: SDI-12 measurement commands
SDI-12 command
1
Values returned
Units
Comments
a
R9!
1. Intensity corrected
accumulation since last
M
command was issued
2. Number of raw bucket tips
since last measurement
3. Total intensity corrected
accumulation since reset
4. Average precipitation intensity
since last measurement
5. Maximum (peak) intensity
since last measurement
1. mm or inch
2. Counts
3. mm or inch
4. mm/hr or in/hr
5. mm/hr or in/hr
Command
does not reset
accumulators
a
I!
014CampbellRain20100SN=
xxx
Where,
xxx
is the serial number
not applicable (NA)
a
V!
1. Hardware Revision starting at 1
2. Reset Code
2 = Power On
11=OS Update
12=Watchdog Timeout
NA
1
a is the sensor address.
8.2 SDI-12 extended commands
Enter the following SDI-12 extended commands using the transparent mode while the computer
is communicating with the data logger through a terminal emulator program. It is accessed
through Campbell Scientific data logger support software or other terminal emulator programs.
Data logger keyboards and displays cannot be used. Refer to
(p. 35) for
more information.
RainVUE-Series SDI-12 Precipitation Sensors