Section 4. Sensor Support
39
Note
The accuracy specification includes only the CR200(X)’s contribution to
measurement error. It does not include the error of sensors.
For example, assume the following (see
Specifications
(p. 35)):
•
Input Voltage: +2000 mV
•
Programmed Measurement Instruction: VoltSE ()
•
CR200(X) Temperature: Between -40°C and +50°C
Accuracy of the measurement is calculated as follows:
Error = Gain Error + Offset Error,
where
Gain Error = ± (2000 * 0.0025)
= ±5 mV
and
Offset Error = 1.2mV
Therefore,
Error = Gain Error + Offset Error
= ±5 mV + 1.2 mV.
= ±6.2 mV
A worst case accuracy in these conditions is
Error = ±(2000 mV * 0.01 + 2.4 mV) = ±22.4 mV
In contrast, the error for a 500 mV input under the same constraints is ±2.45 mV
typical and ±7.4 mV worst case.
Содержание CR200
Страница 32: ...Section 2 Quickstart Tutorial 22 ...
Страница 45: ...Section 3 Overview 35 3 3 Specifications ...
Страница 46: ...Section 3 Overview 36 ...
Страница 58: ...Section 4 Sensor Support 48 ...
Страница 62: ...Section 5 Measurement and Control Peripherals 52 ...
Страница 78: ...Section 8 CR200 X Configuration 68 ...
Страница 102: ...Section 9 Programming 92 ...
Страница 138: ...Section 11 Programming Resource Library 128 ...
Страница 156: ...Section 16 Support Software 146 ...
Страница 160: ...Section 17 Care and Maintenance 150 ...
Страница 166: ...Section 18 Troubleshooting 156 18 3 3 Diagnosis and Fix Procedures 18 3 3 1 Battery Voltage Test ...
Страница 167: ...Section 18 Troubleshooting 157 18 3 3 2 Charging Circuit Test Solar Panel ...
Страница 168: ...Section 18 Troubleshooting 158 18 3 3 3 Charging Circuit Test Transformer ...
Страница 169: ...Section 18 Troubleshooting 159 18 3 3 4 Adjusting Charging Circuit Voltage ...
Страница 170: ...Section 18 Troubleshooting 160 ...
Страница 184: ...Appendix A Glossary 14 ...
Страница 190: ...Appendix B Status Table and Settings 20 ...
Страница 192: ...Appendix C Serial Port Pin Outs 22 ...
Страница 211: ......