CNR1 Net Radiometer
Tilt response:
Max. dev. 2%
Overall ISO classification:
second class
Sensitivity:
10 - 35 µV/(W m
-2
)
Impedance:
125 Ohm nominal
Operating temperature:
-40°C to +80°C
Spectral range:
305-2800 nm (50% points)
Expected signal range for
atmospheric application:
0 - 15 mV typical
Expected accuracy for daily sums:
±
10%
Window heating offset:
Max. 25 W m
-2
at 1000 W m
-2
normal
incidence solar radiation
2.3 CG3 Specifications
Response time 95%:
18 s
Non-stability:
< 1% change per year
Non-linearity:
Max. dev. 2.5% (0-1000 W m
-2
)
Temperature dependence of
sensitivity:
6% (-10°C to +40°C)
Tilt response:
Max. 3% when facing downwards
Field of view:
150 degrees
Sensitivity:
5 - 35 µV/(W m
-2
)
Impedance:
125 Ohm nominal
Operating temperature:
-40°C to +80°C
Temperature range for specified
behavior:
-10°C to +40°C
Measurement range:
-250 to +250 W m
-2
Spectral range:
5 to 50 µm
Expected signal range for
atmospheric application:
-4 to 4 mV
Expected accuracy for daily sums:
±
10%
3. Installation
For measurement of the Net Radiation, it is most important that the instrument
is located in a place that is representative of the entire region that one wishes to
study.
When installed on a mast, the preferred orientation should be such that no
shadow is cast on the Net Radiometer at any time during the day. In the
3